Bone quality assessment around dental implants in cone-beam CT images: effect of rotation mode and metal artefact reduction tool

Lauren Bohner iD, Hian Parize iD, João Victor Cunha Cordeiro iD, Natalia Koerich Laureano iD, Johannes Kleinheinz iD, Ricardo Armini Caldas iD, Dorothea Dagassan-Berndt iD

Dentomaxillofacial Radiology · 2025

DOI: 10.1093/dmfr/twaf003 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Dentomaxillofacial Radiology, 2025. Verfasst von Lauren Bohner und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cranio-Maxillofacial Surgery, University Hospital Muenster, Muenster, 48149, Germany. Hauptthemen laut MEDLINE: Artifacts, Dental Implants, Cone-Beam Computed Tomography. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 36 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: International Team for Implantology.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

The purpose of this study was to evaluate how artefacts caused by titanium and zirconia dental implants affect the bone quality assessment in CBCT images. The effect of scan mode and the use of metal artefact reduction (MAR) algorithm on artefacts suppression were taken in consideration.

Methods

Titanium and zirconia dental implants were installed in porcine bone samples and scanned with two CBCT devices with adjustments on scan mode and with the use of MAR. The control group consisted of bone sample without implant and scanned with full-rotation scan mode without MAR. Artefacts extension and bone quality around implants were measured by deviation of grey values and bone histomorphometry measurements (trabecular volume fraction, bone specific surface, trabecular thickness, and trabecular separation), respectively. Mean difference among groups was assessed by within ANOVA with Bonferroni correction. Correlation between bone quality measurements acquired in the experimental and control groups was assessed by Spearman correlation test (α = .05).

Results

No statistical difference was found for artefacts extension in images acquired by half and full-rotation modes (P = .82). The application of MAR reduced artefacts caused by titanium and zirconia dental implants, showing no statistically significant difference from the control group (titanium: P = .20; zirconia: P = .31). However, there was no correlation between bone quality measurements (P < .05).

Conclusions

Bone quality assessment was affected by the presence of artefacts caused by dental implants. Rotation mode did not affect the appearance of artefacts and bone qualitative measurements. MAR was able to decrease artefacts, however, it did not improve the accuracy of bone quality measurements.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ArtifactsDental ImplantsCone-Beam Computed TomographyAnimalsSwineTitaniumZirconiumAlgorithmsRotationBone Density

Stichwörter der Autor:innen

artefactscone-beam CTdental implants

Substanzen

Dental ImplantsTitaniumZirconiumzirconium oxide

Autor:innen und Institutionen

  • Department of Cranio-Maxillofacial Surgery, University Hospital Muenster, Muenster, 48149, Germany.
  • Department of Dentistry, Federal University of Santa Catarina, Florianópolis, SC 88040-370, 88040370, Brazil.
  • Department of Prosthodontics, School of Dentistry, University of São Paulo, São Paulo, 05508000, Brazil.
  • Department of Oral Pathology, Federal University of Rio Grande do Sul, Porto Alegre, 90650970, Brazil.
  • Center for Dental Imaging, University Center for Dental Medicine Basel UZB, University of Basel, Basel, 4058, Switzerland.

ORCID-Profile: Lauren Bohner, Hian Parize, João Victor Cunha Cordeiro, Natalia Koerich Laureano, Johannes Kleinheinz, Ricardo Armini Caldas, Dorothea Dagassan-Berndt

Förderung

  • International Team for Implantology

Zitieren

APA

Bohner, L., Parize, H., Cordeiro, J. V. C., Laureano, N. K., Kleinheinz, J., Caldas, R. A., & Dagassan-Berndt, D. (2025). Bone quality assessment around dental implants in cone-beam CT images: effect of rotation mode and metal artefact reduction tool. Dentomaxillofacial Radiology. https://doi.org/10.1093/dmfr/twaf003

Vancouver

Bohner L, Parize H, Cordeiro JVC, Laureano NK, Kleinheinz J, Caldas RA, et al. Bone quality assessment around dental implants in cone-beam CT images: effect of rotation mode and metal artefact reduction tool. Dentomaxillofacial Radiology. 2025. doi:10.1093/dmfr/twaf003

BibTeX
@article{Bohner2025,
  author = {Bohner, Lauren and Parize, Hian and Cordeiro, João Victor Cunha and Laureano, Natalia Koerich and Kleinheinz, Johannes and Caldas, Ricardo Armini and Dagassan-Berndt, Dorothea},
  title = {Bone quality assessment around dental implants in cone-beam CT images: effect of rotation mode and metal artefact reduction tool},
  journal = {Dentomaxillofacial Radiology},
  year = {2025},
  doi = {10.1093/dmfr/twaf003}
}
Kostenlos starten →