Vergleichsstudie in der Zeitschrift Clinical Oral Implants Research, online erschienen am 31.05.2024. Verfasst von Thomas Gill und 5 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Dentistry, Barts and the London School of Medicine and Dentistry, Queen Mary University of London, London, UK. Hauptthemen laut MEDLINE: Osseointegration, Dental Implants, Dental Implantation, Endosseous, Dental Prosthesis Design. Die Arbeit wurde bisher 11-mal zitiert (Crossref) und nennt 77 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Institut Straumann AG, Straumann Group.
Abstract (Originalsprache)
Objectives
The aim of the present study was to compare a novel tapered, double‐threaded self‐tapping tissue‐Level design implant (TLC) to a well‐established parallel walled tissue‐level (TL) implant in terms of primary and secondary stability over time. Materials and Methods Test TLC ( n = 10/per timepoint) and control TL ( n = 10/per timepoint) implants were placed in the mandible of minipigs and left for submerged healing for 3, 6, and 12 weeks. Maximum insertion torque and implant stability quotient (ISQ) were measured for each implant at placement. Osseointegration and cortical bone maintenance were histologically evaluated by measuring total bone‐to‐implant contact (BIC) and first bone‐to‐implant contact (fBIC).
Results
A significantly higher maximum insertion torque was measured for the test implant TLC compared to the control TL implant (57.83 ± 24.73 Ncm and 22.62 ± 23.16 Ncm, respectively; p < .001). The mean ISQ values were comparable between the two implant types (75.00 ± 6.70 for TL compared to 75.40 ± 3.20 for TLC, p = .988). BIC was comparable between both implant types at each of the evaluated time points. The fBIC was found to be significantly more coronal at 12 weeks for the TLC implant compared to the TL implant (0.31 ± 0.83 mm for TLC compared to −0.22 ± 0.85 for TL, p = .027).
Conclusion
The novel tapered tissue level design implant showed improved primary stability and an overall improved crestal bone height maintenance compared to the parallel walled design at 12 weeks.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Institute of Dentistry, Barts and the London School of Medicine and Dentistry, Queen Mary University of London, London, UK.
- Department of Oral Surgery, University Center for Dental Medicine, University of Basel, Basel, Switzerland.
- Department of Periodontology, School of Dental Medicine, University of Bern, Bern, Switzerland.
- Department of Preclinical Research, Institut Straumann AG, Basel, Switzerland.
- Faculty of Medicine and Health Technology, University of Tampere, Tampere, Finland.
- Department of Restorative Dentistry, the Royal London Dental Hospital, Barts Health NHS Trust, London, UK.
ORCID-Profile: Thomas Gill, Sebastian Kühl, Benjamin Pippenger, Shakeel Shahdad
Förderung
- Institut Straumann AG
- Straumann Group
Zitieren
APA
Gill, T., Kühl, S., Rawlinson, S., Pippenger, B., Bellon, B., & Shahdad, S. (2024). Primary stability and osseointegration comparing a novel tapered design tissue‐level implant with a parallel design tissue‐level implant. An experimental in vivo study. Clinical Oral Implants Research, 35(9), 1114-1127. https://doi.org/10.1111/clr.14301
Vancouver
Gill T, Kühl S, Rawlinson S, Pippenger B, Bellon B, Shahdad S. Primary stability and osseointegration comparing a novel tapered design tissue‐level implant with a parallel design tissue‐level implant. An experimental in vivo study. Clinical Oral Implants Research. 2024;35(9):1114-1127. doi:10.1111/clr.14301
BibTeX
@article{Gill2024,
author = {Gill, Thomas and Kühl, Sebastian and Rawlinson, Simon and Pippenger, Benjamin and Bellon, Benjamin and Shahdad, Shakeel},
title = {Primary stability and osseointegration comparing a novel tapered design tissue‐level implant with a parallel design tissue‐level implant. An experimental in vivo study},
journal = {Clinical Oral Implants Research},
year = {2024},
volume = {35},
number = {9},
pages = {1114--1127},
doi = {10.1111/clr.14301}
}
EUVIMED