Comparative osseointegration of hydrophobic tissue‐level tapered implants—A preclinical in vivo study
Clinical Oral Implants Research · 2024
DOI: 10.1111/clr.14318 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Clinical Oral Implants Research, 2024. Verfasst von Jean-Claude Imber und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Periodontology, School of Dental Medicine, University of Bern, Bern, Switzerland. Hauptthemen laut MEDLINE: Osseointegration, Dental Implants, Dental Implantation, Endosseous, Swine, Miniature. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 48 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: Institut Straumann AG, Straumann Group.
Abstract (Originalsprache)
Purpose
To histometrically compare the osseointegration and crestal bone healing of a novel tapered, self‐cutting tissue‐level test implant with a standard tissue‐level control implant in a submerged healing regimen. Materials and Methods In a mandibular minipig model, implants were inserted and evaluated histometrically after a healing period of 3, 6, and 12 weeks. The primary outcome was the evaluation of bone‐to‐implant contact (BIC) and secondary outcomes were primary stability as per insertion torque and first BIC (fBIC). Outcomes for the test and control implants were compared using Wilcoxon signed‐rank tests and mixed linear regression models.
Results
Insertion torque values were significantly higher for the test (50.0 ± 26.4 Ncm) compared to the control implants (35.2 ± 19.7 Ncm, p = .0071). BIC values of test implants were non‐inferior to those of control implants over the investigated study period. After 12 weeks, the corresponding values measured were 81.62 ± 11.12% and 90.41 ± 4.81% ( p = .1763) for test and control implants, respectively. Similarly, no statistical difference was found for fBIC values, except for the 12 weeks outcome that showed statistically lower values for the test (−675.58 ± 590.88 μm) compared to control implants (−182.75 ± 197.40 μm, p = .0068).
Conclusions
Novel self‐cutting tissue‐level implants demonstrated noninferior osseointegration and crestal bone height maintenance to the tissue‐level implants. Histometric outcomes between both implants demonstrated test implants were statistically noninferior to control implants, despite substantial differences in the bone engagement mechanism and resulting differences in insertion torque and qualitative bone healing patterns.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Periodontology, School of Dental Medicine, University of Bern, Bern, Switzerland.
- Department of Preclinical Research, Institut Straumann AG, Basel, Switzerland.
ORCID-Profile: Jean-Claude Imber, Andrea Roccuzzo, Dieter D Bosshardt, Anton Sculean, Benjamin E Pippenger
Förderung
- Institut Straumann AG
- Straumann Group
Zitieren
APA
Imber, J. C., Khandanpour, A., Roccuzzo, A., Irani, D. R., Bosshardt, D. D., Sculean, A., & Pippenger, B. E. (2024). Comparative osseointegration of hydrophobic tissue‐level tapered implants—A preclinical in vivo study. Clinical Oral Implants Research. https://doi.org/10.1111/clr.14318
Vancouver
Imber JC, Khandanpour A, Roccuzzo A, Irani DR, Bosshardt DD, Sculean A, et al. Comparative osseointegration of hydrophobic tissue‐level tapered implants—A preclinical in vivo study. Clinical Oral Implants Research. 2024. doi:10.1111/clr.14318
BibTeX
@article{Imber2024,
author = {Imber, Jean-Claude and Khandanpour, Azita and Roccuzzo, Andrea and Irani, Delia R and Bosshardt, Dieter D and Sculean, Anton and Pippenger, Benjamin E},
title = {Comparative osseointegration of hydrophobic tissue‐level tapered implants—A preclinical in vivo study},
journal = {Clinical Oral Implants Research},
year = {2024},
doi = {10.1111/clr.14318}
}
EUVIMED