Autoclaving‐induced dimensional changes of three‐dimensional printed surgical guides: An in vitro study
Clinical Oral Implants Research · 2024-08 · 35 (8) · 821-829
DOI: 10.1111/clr.14158 ↗Kurz gefasst
Fachartikel in der Zeitschrift Clinical Oral Implants Research, online erschienen am 10.08.2023. Verfasst von Mira Hüfner und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthodontics, University Hospital of Düsseldorf, Düsseldorf, Germany. Hauptthemen laut MEDLINE: Printing, Three-Dimensional. Die Arbeit wurde bisher 17-mal zitiert (Crossref) und nennt 36 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: 3D Agency oHG, Dentalstudio Sankt Augustin GmbH.
Abstract (Originalsprache)
Objectives
Surgical guides are frequently used for dental implant placement. The aim of this study was to evaluate the impact of the 3D printing process itself and subsequent steam autoclaving on the dimensional stability of five different resin/printer combinations (RPCs). Materials and Methods Fifty identical surgical guides (10 per group) were produced consisting of five RPCs. Half of the guides (5 per group) were steam autoclaved with cycle 1 (121°C, 1 bar, 20.5 min) and the other half with cycle 2 (134°C, 2 bar, 5.5 min). All guides were scanned with a structured‐light (SL) 3D scanner before (T0) and after (T1) autoclaving. Linear measurements along the x‐, y‐, and z‐axes were performed at landmarks on the original STL file and on SL scans at T0 and T1, respectively. Wilcoxon signed‐rank test, Kruskal–Wallis test, and linear mixed‐effects models were performed, depending on the analysis.
Results
Three‐dimensional printing was associated with significant dimensional alterations for all RPCs. Steam autoclaving using cycle 1 was associated with significant shrinkage in x‐ (1 RPC), y‐ (2 RPCs), and z‐direction (2 RPCs), while cycle 2 was also associated with shrinkage in x‐ (2 RPCs), y‐ (1 RPC), and z‐direction (1 RPC). One resin did not present any dimensional changes independently of the cycle.
Conclusions
The majority of the guides presented minor but significant shrinkage due to 3D printing itself and both steam autoclaving cycles, the extent varied between different RPCs. Whether these changes compromise implant placement accuracy remains to be investigated.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Orthodontics, University Hospital of Düsseldorf, Düsseldorf, Germany.
- Department of Oral Surgery, University Hospital of Düsseldorf, Düsseldorf, Germany.
- Department of Neurosciences, University of Padua, Padua, Italy.
- Oral Technology, University Hospital Bonn, Bonn, Germany.
ORCID-Profile: Mira Hüfner, Giulia Brunello, Kathrin Becker
Förderung
- 3D Agency oHG
- Dentalstudio Sankt Augustin GmbH
Zitieren
APA
Hüfner, M., David, S., Brunello, G., Kerberger, R., Rauch, N., Busch, C. V., Drescher, D., Bourauel, C., & Becker, K. (2024). Autoclaving‐induced dimensional changes of three‐dimensional printed surgical guides: An in vitro study. Clinical Oral Implants Research, 35(8), 821-829. https://doi.org/10.1111/clr.14158
Vancouver
Hüfner M, David S, Brunello G, Kerberger R, Rauch N, Busch CV, et al. Autoclaving‐induced dimensional changes of three‐dimensional printed surgical guides: An in vitro study. Clinical Oral Implants Research. 2024;35(8):821-829. doi:10.1111/clr.14158
BibTeX
@article{Hfner2024,
author = {Hüfner, Mira and David, Samuel and Brunello, Giulia and Kerberger, Robert and Rauch, Nicole and Busch, Caroline Viola and Drescher, Dieter and Bourauel, Christoph and Becker, Kathrin},
title = {Autoclaving‐induced dimensional changes of three‐dimensional printed surgical guides: An in vitro study},
journal = {Clinical Oral Implants Research},
year = {2024},
volume = {35},
number = {8},
pages = {821--829},
doi = {10.1111/clr.14158}
}
EUVIMED