Impact of Autoclaving on the Dimensional Stability of 3D‐Printed Guides for Orthodontic Mini‐Implant Insertion – An In Vitro Study
Clinical and Experimental Dental Research · 2025
DOI: 10.1002/cre2.70111 ↗Kurz gefasst
Fachartikel in der Zeitschrift Clinical and Experimental Dental Research, 2025. Verfasst von Samuel David und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthodontics, University Hospital of Düsseldorf, Düsseldorf, Germany. Hauptthemen laut MEDLINE: Printing, Three-Dimensional, Sterilization, Orthodontic Anchorage Procedures, Dental Implants. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 49 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: The authors received no specific funding for this work..
Abstract (Originalsprache)
Objectives
This study aimed to assess the effect of the printing process itself and steam autoclaving on the geometrical stability of 3D‐printed guides for mini‐implant insertion. Material and Methods Fifty guides ( n = 10 per group) were printed with five printer/resin combinations (PRCs) from the same STL file using either digital light processing (DLP/EG, DLP/Next, DLP/Opti), desktop stereolithography (SLA/DSG) or liquid crystal display stereolithography printers (LCD/Amber). Half were sterilized by steam autoclaving with Cycle 1 (121°C, 1 bar, 20.5 min), half with Cycle 2 (134°C, 2 bars, 5.5 min). Before (T0) and after sterilization (T1) the guides were scanned with a structured light 3D scanner, and selected guides also with micro‐CT for validation. Linear measurements were performed in three axes on STL, and on T0 and T1 scans. Linear mixed‐effects models were used, followed by post‐hoc tests in case of significance.
Results
Measurements at T0 and T1 differed significantly from STL in both x ‐ and y ‐axis (4 and 3 PRCs, respectively) ( p < 0.05); in z ‐axis only DLP/Next showed significant differences between T0 and STL ( p < 0.001). The comparison between T0 and T1 revealed significant differences in x ‐axis for DLP/Next and DLP/Opti after Cycle 1 and Cycle 2, respectively ( p < 0.05), while in the y ‐axis no intra‐group difference was recorded. In the z ‐axis all PRCs except for SLA/DSG exhibited significant shrinkage (for Cycles 1 or 2). Differences between the two cycles at T1 were registered only in z ‐axis (DLP/Next and LCD/Amber).
Conclusions
Compared with the original, all PRCs except for SLA/DSG presented significant changes in their dimensional stability owing to the printing process itself and/or the sterilization. If these changes are of clinical significance, they remain to be verified. Clinical Relevance With the utilized design, the guides fabricated with SLA provided lower dimensional changes as compared to the ones produced by the other printing techniques.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Orthodontics, University Hospital of Düsseldorf, Düsseldorf, Germany.
- Department of Oral Surgery, University of Düsseldorf, Düsseldorf, Germany.
- Department of Orthodontics and Dentofacial Orthopaedics, Charité - Universitätsmedizin, Freie Universität Berlin and Humboldt-Universität zu Berlin, Berlin, Germany.
- Department of Neurosciences, School of Dentistry, University of Padua, Padua, Italy.
Förderung
- The authors received no specific funding for this work.
Zitieren
APA
David, S., Hüfner, M., Rauch, N., Kerberger, R., Drescher, D., Brunello, G., & Becker, K. (2025). Impact of Autoclaving on the Dimensional Stability of 3D‐Printed Guides for Orthodontic Mini‐Implant Insertion – An In Vitro Study. Clinical and Experimental Dental Research. https://doi.org/10.1002/cre2.70111
Vancouver
David S, Hüfner M, Rauch N, Kerberger R, Drescher D, Brunello G, et al. Impact of Autoclaving on the Dimensional Stability of 3D‐Printed Guides for Orthodontic Mini‐Implant Insertion – An In Vitro Study. Clinical and Experimental Dental Research. 2025. doi:10.1002/cre2.70111
BibTeX
@article{David2025,
author = {David, Samuel and Hüfner, Mira and Rauch, Nicole and Kerberger, Robert and Drescher, Dieter and Brunello, Giulia and Becker, Kathrin},
title = {Impact of Autoclaving on the Dimensional Stability of 3D‐Printed Guides for Orthodontic Mini‐Implant Insertion – An In Vitro Study},
journal = {Clinical and Experimental Dental Research},
year = {2025},
doi = {10.1002/cre2.70111}
}
EUVIMED