Effects of sterilization and disinfection methods on digitally designed surgical implant guide accuracy: An in vitro study

Ruikun Li, Jie Zhu, Shuo Wang, Xueyuan Li, Songhang Li

Clinical Implant Dentistry and Related Research · 2024

DOI: 10.1111/cid.13350 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Clinical Implant Dentistry and Related Research, 2024. Verfasst von Ruikun Li und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Implant Dentistry, West China Hospital of Stomatology, Sichuan University, Chengdu, China. Hauptthemen laut MEDLINE: Sterilization, Disinfection. Die Arbeit wurde bisher 10-mal zitiert (Crossref) und nennt 31 Literaturangaben. Gefördert von: Postdoctoral Science Foundation of China, Postdoctoral Foundation of Sichuan University, Miaozi Project in Science and Technology Innovation Program of Sichuan Province.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Surgical guides are commonly used to assist with dental implant placement. This study investigated the effects of five sterilization and disinfection methods on the accuracy of implant guides.

Methods

Thirty surgical guides (five in each group) were designed and printed (with digital light processing technology) using different sterilization or disinfection methods categorized into six groups: hydrogen peroxide sterilization (group one); glutaraldehyde sterilization (group two); autoclaving (group three); plasma sterilization (group four); iodophor disinfection (group five); and blank group (group six). Verification was determined using three methods: distance and angle between the cross‐shaped marks, deformation after superimposing the guides, and displacement and axial changes in the virtual implant.

Results

After disinfection and sterilization, the guides in the autoclaving and iodophor groups showed a more pronounced color change and the guide in the autoclaving group had visible cracks. More significant changes were observed in the H 2 O 2 , glutaraldehyde, autoclaving, and iodophor groups regarding deformation after superimposing the guides and the distance and angle between the cross‐shaped marks. The average labial deformation values (mm) of the first through fifth groups of guides were 0.283, 0.172, 0.289, 0.153, and 0.188, respectively. All groups were statistically different from the blank group for displacement and axial changes of the virtual implant ( p < 0.05).

Conclusion

The sizes of almost all surgical guides changed after sterilization and disinfection treatments, with between‐group differences. Plasma sterilization was more suitable for surgical guide sterilization because of the smaller deformations after treatment.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

SterilizationDisinfectionHumansIn Vitro TechniquesHydrogen PeroxideSurgery, Computer-AssistedGlutaralDental ImplantsComputer-Aided DesignDental Implantation, Endosseous

Stichwörter der Autor:innen

accuracycomputer‐aideddental implantsdisinfectionsterilizationsurgical guides

Substanzen

Hydrogen PeroxideGlutaralDental Implants

Autor:innen und Institutionen

  • Department of Implant Dentistry, West China Hospital of Stomatology, Sichuan University, Chengdu, China.
  • State Key Laboratory of Oral Diseases, National Clinical Research Center for Oral Diseases, West China Hospital of Stomatology, Sichuan University, Chengdu, China.
  • Department of Ophthalmology, the General Hospital of Western Theater Command, Chengdu, China.

Förderung

  • Postdoctoral Science Foundation of China
  • Postdoctoral Foundation of Sichuan University
  • Miaozi Project in Science and Technology Innovation Program of Sichuan Province

Zitieren

APA

Li, R., Zhu, J., Wang, S., Li, X., & Li, S. (2024). Effects of sterilization and disinfection methods on digitally designed surgical implant guide accuracy: An in vitro study. Clinical Implant Dentistry and Related Research. https://doi.org/10.1111/cid.13350

Vancouver

Li R, Zhu J, Wang S, Li X, Li S. Effects of sterilization and disinfection methods on digitally designed surgical implant guide accuracy: An in vitro study. Clinical Implant Dentistry and Related Research. 2024. doi:10.1111/cid.13350

BibTeX
@article{Li2024,
  author = {Li, Ruikun and Zhu, Jie and Wang, Shuo and Li, Xueyuan and Li, Songhang},
  title = {Effects of sterilization and disinfection methods on digitally designed surgical implant guide accuracy: An in vitro study},
  journal = {Clinical Implant Dentistry and Related Research},
  year = {2024},
  doi = {10.1111/cid.13350}
}
Kostenlos starten →