Influence of straight versus angulated screw channel titanium bases on failure loads of two‐piece ceramic and titanium implants restored with screw‐retained monolithic crowns: An in‐vitro study

E Helal, P C Gierthmuehlen iD, E A Bonfante iD, T M B Campos, L S Prott iD, R Langner iD, F A Spitznagel iD

Clinical Oral Implants Research · 2023

DOI: 10.1111/clr.14157 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Clinical Oral Implants Research, 2023. Verfasst von E Helal und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Prosthodontics, Medical Faculty and University Hospital Düsseldorf, Heinrich-Heine-University Düsseldorf, Düsseldorf, Germany. Hauptthemen laut MEDLINE: Zirconium, Dental Implants. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 55 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

To analyze the influence of titanium‐base (straight [SSC]/angulated‐screw‐channel [ASC]) on failure‐loads and bending‐moments of two‐piece ceramic and titanium–zirconium implants restored with monolithic‐zirconia crowns after fatigue. Materials and Methods Thirty‐two anterior monolithic‐screw‐retained zirconia crowns were divided into four groups ( n = 8/group) according to the factors: (1) type of implant material: two‐piece titanium–zirconium implant (Ti–Zr; control‐group) versus two‐piece ceramic implant (CI; test‐group) and (2) type of titanium‐base: SSC (0° angle) versus ASC (25°). An intact implant was used for field emission gun‐scanning electronic microscopy (FEG‐SEM) characterization and Raman spectroscopy for phase analyses and residual stress quantification. All samples were exposed to fatigue with thermodynamic loading (1.2‐million‐cycles, 49 N, 1.6 Hz, 5–55°C) at a 30° angle. Surviving specimens were loaded until failure (SLF) and bending moments were recorded. Failed samples were examined using light microscope and SEM. Statistical analyses included ANOVA and Mann–Whitney U ‐test.

Results

Raman‐spectroscopy revealed the presence of residual compressive stresses. FEG‐SEM revealed a roughened surface between threads and polished surface at the cervical‐collar of the ceramic implant. All samples survived fatigue and were free of complications. Mean bending‐moments (±SD) were: Ti‐Zr‐0: 241 ± 45 N cm, Ti‐Zr‐25: 303 ± 86 N cm, CI‐0: 326 ± 58 N cm, CI‐25: 434 ± 71 N cm. Titanium‐base and implant‐material had significant effects in favor of ASC titanium bases ( p = .001) and ceramic‐implants ( p < .001). Failure analysis after SLF revealed severe fractures in ceramic implants, whereas titanium implants were restricted to plastic deformation.

Conclusions

Ceramic and titanium implants exhibited high reliability after fatigue, with no failures. From a mechanical perspective, titanium bases with ASC can be recommended for both ceramic and titanium implants and are safe for clinical application.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ZirconiumDental ImplantsTitaniumReproducibility of ResultsMaterials TestingDental AbutmentsDental Implant-Abutment DesignDental Restoration FailureCeramicsCrownsBone ScrewsAlloys

Stichwörter der Autor:innen

agingceramicsdental implantfatiguematerial testingzirconia

Substanzen

ZirconiumTitaniumDental Implantszirconium oxideTi-Zr alloyAlloys

Autor:innen und Institutionen

  • Department of Prosthodontics, Medical Faculty and University Hospital Düsseldorf, Heinrich-Heine-University Düsseldorf, Düsseldorf, Germany.
  • Department of Prosthodontics and Periodontology, Bauru School of Dentistry, University of Sao Paulo, Bauru, Brazil.
  • Institute of Systems Neuroscience, Medical Faculty, Heinrich-Heine-University Düsseldorf, Düsseldorf, Germany.
  • Institute of Neuroscience and Medicine, Brain and Behavior (INM-7), Research Center Jülich, Jülich, Germany.

ORCID-Profile: P C Gierthmuehlen, E A Bonfante, L S Prott, R Langner, F A Spitznagel

Zitieren

APA

Helal, E., Gierthmuehlen, P. C., Bonfante, E. A., Campos, T. M. B., Prott, L. S., Langner, R., & Spitznagel, F. A. (2023). Influence of straight versus angulated screw channel titanium bases on failure loads of two‐piece ceramic and titanium implants restored with screw‐retained monolithic crowns: An in‐vitro study. Clinical Oral Implants Research. https://doi.org/10.1111/clr.14157

Vancouver

Helal E, Gierthmuehlen PC, Bonfante EA, Campos TMB, Prott LS, Langner R, et al. Influence of straight versus angulated screw channel titanium bases on failure loads of two‐piece ceramic and titanium implants restored with screw‐retained monolithic crowns: An in‐vitro study. Clinical Oral Implants Research. 2023. doi:10.1111/clr.14157

BibTeX
@article{Helal2023,
  author = {Helal, E and Gierthmuehlen, P C and Bonfante, E A and Campos, T M B and Prott, L S and Langner, R and Spitznagel, F A},
  title = {Influence of straight versus angulated screw channel titanium bases on failure loads of two‐piece ceramic and titanium implants restored with screw‐retained monolithic crowns: An in‐vitro study},
  journal = {Clinical Oral Implants Research},
  year = {2023},
  doi = {10.1111/clr.14157}
}
Kostenlos starten →