Vergleichsstudie in der Zeitschrift Clinical Implant Dentistry and Related Research, online erschienen am 17.11.2026. Verfasst von Xu Zhao und 5 weiteren Autor:innen. Beteiligte Einrichtung: First Clinical Division, Peking University School and Hospital for Stomatology & National Clinical Research Center for Oral Diseases & National Engin…. Hauptthemen laut MEDLINE: Dental Implantation, Endosseous, Jaw, Edentulous, Partially, Surgery, Computer-Assisted, Robotic Surgical Procedures. Die Arbeit nennt 32 Literaturangaben. Gefördert von: Program for New Clinical Techniques and Therapies of Peking University School and Hospital of Stomatology, Beijing Natural Science Foundation, Natural Science Foundation of Beijing Municipality.
Abstract (Originalsprache)
Aim
To evaluate the accuracy and chair‐side efficiency of static (s‐CAIS), dynamic (d‐CAIS), and autonomous robotic (r‐CAIS) computer‐assisted implant surgery in healed partially edentulous sites. Materials and Methods In this retrospective comparative study, 73 patients with 92 implants, treated from September 2022 to October 2024, were included (s‐CAIS: 21 patients/33 implants; d‐CAIS: 29 patients/32 implants; r‐CAIS: 23 patients/27 implants). Pre‐ and postoperative CBCT scans were used to assess linear and angular deviations. The duration of chair‐side preparation and drilling/implant placement was recorded for d‐CAIS and r‐CAIS.
Results
At implant level, mean platform deviations were 0.85 ± 0.38 mm (s‐CAIS), 1.37 ± 0.71 mm (d‐CAIS), and 0.59 ± 0.32 mm (r‐CAIS; median [IQR]: 0.54 [0.30] mm); apex deviations were 1.36 ± 0.51, 1.48 ± 0.66, and 0.64 ± 0.52 mm (r‐CAIS; median [IQR]: 0.53 [0.30] mm); angular deviations were 4.34° ± 2.13°, 2.53° ± 0.99°, and 1.54° ± 1.50° (r‐CAIS; median [IQR]: 1.28° [1.40°]). At patient level, platform deviations were 0.88 ± 0.35, 1.37 ± 0.69, and 0.57 ± 0.29 mm; apex deviations 1.38 ± 0.44, 1.48 ± 0.65, and 0.58 ± 0.43 mm; angular deviations 4.48° ± 1.81°, 2.52° ± 0.97°, and 1.45° ± 1.30°. r‐CAIS showed significantly superior accuracy compared with s‐CAIS and d‐CAIS ( p < 0.05). Because r‐CAIS data were not normally distributed, medians [IQR] were additionally calculated and are reported parenthetically. Drilling and placement time was longer for d‐CAIS than r‐CAIS (18.06 ± 5.54 vs. 10.81 ± 3.55 min, p < 0.05), while chair‐side preparatory time did not differ (10.78 ± 2.53 vs. 9.61 ± 2.72 min, p > 0.05).
Conclusions
Among the three systems, r‐CAIS provided the highest accuracy and was more time‐efficient during the intra‐operative (chair‐side) phase than d‐CAIS in partially edentulous patients. Clinical Relevance In patients with partial edentulism, r‐CAIS demonstrated superior accuracy compared with both s‐CAIS and d‐CAIS approaches. Furthermore, r‐CAIS offered greater chair‐side efficiency during the intra‐operative phase than d‐CAIS. This cutting‐edge technology holds great promise for enhancing the precision of dental implant procedures, offering substantial benefits to both surgeons and patients.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- First Clinical Division, Peking University School and Hospital for Stomatology & National Clinical Research Center for Oral Diseases & National Engineering Laboratory for Digital and Material Technology for Stomatology,…
- Division of Conservative Dentistry and Periodontology, Center of Clinical Research, Dental School, Medical University of Vienna, Vienna, Austria.
ORCID-Profile: Xu Zhao, MiaoZhen Wang, Tao Yu, Feng Liu
Förderung
- Program for New Clinical Techniques and Therapies of Peking University School and Hospital of Stomatology
- Beijing Natural Science Foundation
- Natural Science Foundation of Beijing Municipality
Zitieren
APA
Zhao, X., Wang, M., Yu, T., Ma, F., Rausch-Fan, X., & Liu, F. (2026). Static, Dynamic and Robotic CAIS in Healed Partially Edentulous Sites: Accuracy and Chair‐Side Time. Clinical Implant Dentistry and Related Research, 28(6). https://doi.org/10.1111/cid.70201
Vancouver
Zhao X, Wang M, Yu T, Ma F, Rausch-Fan X, Liu F. Static, Dynamic and Robotic CAIS in Healed Partially Edentulous Sites: Accuracy and Chair‐Side Time. Clinical Implant Dentistry and Related Research. 2026;28(6). doi:10.1111/cid.70201
BibTeX
@article{Zhao2026,
author = {Zhao, Xu and Wang, MiaoZhen and Yu, Tao and Ma, FeiFei and Rausch-Fan, Xiaohui and Liu, Feng},
title = {Static, Dynamic and Robotic CAIS in Healed Partially Edentulous Sites: Accuracy and Chair‐Side Time},
journal = {Clinical Implant Dentistry and Related Research},
year = {2026},
volume = {28},
number = {6},
doi = {10.1111/cid.70201}
}
EUVIMED