EUVIMEDObjevte EUVIMED
Crossref

Effects of Three‐Dimensional Print Offsets of Sleeves on Implant Placement Trueness: An In Vitro Study

Chenli Kang, Simeng Hao, Chunlei Zhang, Xueyuan Li iD, Songhang Li iD

Clinical Implant Dentistry and Related Research · 2025

DOI: 10.1111/cid.70101 ↗

At a glance

Journal article in Clinical Implant Dentistry and Related Research, 2025. Written by Chenli Kang and 4 co-authors. Institution: Department of Oral Implantology, West China Hospital of Stomatology, Sichuan University, Chengdu, China. Major MEDLINE topics: Printing, Three-Dimensional, Dental Implantation, Endosseous, Surgery, Computer-Assisted, Dental Implants. Cited 2 times so far (Crossref); 30 references. Funded by: National Natural Science Foundation of China, Natural Science Foundation of Sichuan Province, Postdoctoral Science Foundation of China.

AbstraktSubjectsBibliografické údajeCiteRelated studies

Abstract (original language)

Introduction

This study assesses the effects of different three‐dimensional (3D) print offsets of sleeves on sleeve and implant placement trueness in guided implant surgery.

Methods

Twenty‐five sleeves were installed on 3D‐printed surgical guides, with sleeve offsets of 0.02, 0.04, 0.06, 0.08, and 0.10 mm (five groups with five sleeves per group). The actual surgical guides were scanned to generate digital files, which were compared with the reference group based on design positions. The experimental verification and evaluation method involves assessing sleeve installations and measuring the deviations between the actual position of the sleeve and implant compared with the reference group.

Results

The sleeve installations in the 0.02‐mm offset group were not acceptable, as there were visible cracks in the sleeve; in the case of the 0.04‐mm offset group, there were also installation difficulties. At the sleeve position level, there were larger deviations in the 0.10‐mm offset group, whereas the 0.08‐mm group had smaller deviations in the labial and palatal directions ( p < 0.05). At the implant level, the 0.10‐mm offset group showed a significantly larger deviation (labial: 0.453 ± 0.165 mm; palatal: 0.399 ± 0.160 mm; p < 0.05). As for the mesial–distal direction, there were no significant differences between the groups.

Conclusions

Print offsets of sleeves had a significant effect on the ease of installation and the placement accuracy of sleeves and implants. Under the printer and sleeves conditions used in this experiment, smaller offsets (0.02 and 0.04 mm) caused installation difficulties and misalignments. Larger offsets (0.10 mm) created noticeable deviations. Sleeve offsets of 0.06 and 0.08 mm provided a balance between ease of installation and high placement trueness, making them more suitable in this experiment. Excessive or insufficient print offsets of sleeves can affect installation conditions and positional trueness of sleeves and implants.

Subjects

MeSH terms (MEDLINE)

Printing, Three-DimensionalDental Implantation, EndosseousSurgery, Computer-AssistedDental ImplantsHumansIn Vitro Techniques

Author keywords

accuracycomputer‐aidedimplantsurgical guidesthree‐dimensional

Substances

Dental Implants

Authors and institutions

  • Department of Oral Implantology, West China Hospital of Stomatology, Sichuan University, Chengdu, China.
  • State Key Laboratory of Oral Diseases, National Clinical Research Center for Oral Diseases, West China Hospital of Stomatology, Sichuan University, Chengdu, China.
  • School and Hospital of Stomatology, China Medical University, Shenyang, China.
  • Department of Ophthalmology, The General Hospital of Western Theater Command, Chengdu, China.

ORCID profiles: Xueyuan Li, Songhang Li

Funding

  • National Natural Science Foundation of China
  • Natural Science Foundation of Sichuan Province
  • Postdoctoral Science Foundation of China
  • Postdoctoral Fellowship Program of CPSF
  • Sichuan University Interdisciplinary Innovation Fund
  • Research and Development Program, West China Hospital of Stomatology Sichuan University
  • Young Innovation Project of Sichuan Medical Association

Cite

APA

Kang, C., Hao, S., Zhang, C., Li, X., & Li, S. (2025). Effects of Three‐Dimensional Print Offsets of Sleeves on Implant Placement Trueness: An In Vitro Study. Clinical Implant Dentistry and Related Research. https://doi.org/10.1111/cid.70101

Vancouver

Kang C, Hao S, Zhang C, Li X, Li S. Effects of Three‐Dimensional Print Offsets of Sleeves on Implant Placement Trueness: An In Vitro Study. Clinical Implant Dentistry and Related Research. 2025. doi:10.1111/cid.70101

BibTeX
@article{Kang2025,
  author = {Kang, Chenli and Hao, Simeng and Zhang, Chunlei and Li, Xueyuan and Li, Songhang},
  title = {Effects of Three‐Dimensional Print Offsets of Sleeves on Implant Placement Trueness: An In Vitro Study},
  journal = {Clinical Implant Dentistry and Related Research},
  year = {2025},
  doi = {10.1111/cid.70101}
}
Start for free →