In Vitro Analysis of Structural Integrity and Surface Alterations of Reused Healing Abutments
Clinical Implant Dentistry and Related Research · 2025
DOI: 10.1111/cid.70064 ↗Kurz gefasst
Fachartikel in der Zeitschrift Clinical Implant Dentistry and Related Research, 2025. Verfasst von Won-Woo Lee und 5 weiteren Autor:innen. Beteiligte Einrichtung: Division of Oral and Maxillofacial Surgery, Department of Dentistry, College of Medicine, Hanyang University, Seoul, Republic of Korea. Hauptthemen laut MEDLINE: Dental Abutments. Die Arbeit nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
Purpose
This study aimed to investigate the changes in healing abutments (HAs) after use based on an in vitro comparison with unused HAs and to evaluate the effectiveness and clinical implications of reusing HAs. Materials and Methods Fifty used HAs were collected from affiliated clinics of the Department of Dentistry at Hanyang University Hospital and analyzed for surface damage, protein contamination, and microgap formation using three‐dimensional laser microscopy and microcomputed tomography. The interfacial microgap between the implant and the abutment was measured at different tightening torques (5 N·cm, 10 N·cm, and 15 N·cm). Additionally, bacterial leakage and growth at various tightening torques were assessed by incubating reused HAs over different time periods.
Results
Reused HAs exhibited significant surface roughness, protein contamination, and larger microgap compared to unused HAs. The average microgap for reused HAs was 43 μm, whereas unused HAs showed no detectable gap. Bacterial leakage was significantly higher with reused HAs, particularly in those subjected to more than two tightening cycles. Tightening torques of 15 N·cm effectively eliminated the microgap and minimized bacterial leakage, whereas lower torques (5 N·cm and 10 N·cm) resulted in considerable bacterial growth.
Conclusion
Reusing HAs increases the risk of surface damage, protein contamination, microgap formation, and bacterial leakage, potentially compromising implant treatment outcomes. Higher tightening torque (15 N·cm) significantly reduces microgap and bacterial leakage at the implant ‐ HA interface. Clinicians are advised to limit the reuse of HAs. However, if reuse is necessary, an appropriate tightening torque should be applied following a careful assessment of the clinical conditions of each HA.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Division of Oral and Maxillofacial Surgery, Department of Dentistry, College of Medicine, Hanyang University, Seoul, Republic of Korea.
- Department of Medicine, College of Medicine, Hanyang University, Seoul, Korea.
- Biostatistical Consulting and Research Lab, Medical Research Collaborating Center, Hanyang University, Seoul, Korea.
ORCID-Profile: Kyung-Gyun Hwang, Chang-Joo Park
Zitieren
APA
Lee, W. W., Jun, J. H., Lee, J. H., Ryu, S., Hwang, K. G., & Park, C. J. (2025). In Vitro Analysis of Structural Integrity and Surface Alterations of Reused Healing Abutments. Clinical Implant Dentistry and Related Research. https://doi.org/10.1111/cid.70064
Vancouver
Lee WW, Jun JH, Lee JH, Ryu S, Hwang KG, Park CJ. In Vitro Analysis of Structural Integrity and Surface Alterations of Reused Healing Abutments. Clinical Implant Dentistry and Related Research. 2025. doi:10.1111/cid.70064
BibTeX
@article{Lee2025,
author = {Lee, Won-Woo and Jun, Jong-Hun and Lee, Jeong-Hun and Ryu, Soorack and Hwang, Kyung-Gyun and Park, Chang-Joo},
title = {In Vitro Analysis of Structural Integrity and Surface Alterations of Reused Healing Abutments},
journal = {Clinical Implant Dentistry and Related Research},
year = {2025},
doi = {10.1111/cid.70064}
}
EUVIMED