Vergleichsstudie in der Zeitschrift Clinical Implant Dentistry and Related Research, 2025. Verfasst von Gustavo Harfagar und 5 weiteren Autor:innen. Beteiligte Einrichtung: Private Practice at Matanzas Clinic, Navidad, Chile. Hauptthemen laut MEDLINE: Jaw, Edentulous, Dental Impression Technique, Dental Implants, Computer-Aided Design. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 63 Literaturangaben. Gefördert von: Institute Straumann AG.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Aim

To compare the linear and angular deviations of conventional implant (CI) and digital implant (DI) impression techniques in edentulous jaws with four or six implants. Materials and Methods Twenty participants (12 men, 8 women; mean age 58.6 years) with complete edentulous maxillary ( n = 8) or mandibular ( n = 12) arches were included. Each patient received four or six dental implants (Straumann BLX). Both CI and DI were performed using randomized sequences. Linear and angular deviations were measured between the reference scan (coordinated measuring machine) and the CI (desktop scanner) and DI (intraoral scanner, IOS) using CATIA software (Dassault Systèmes). Framework passivity was evaluated using the Sheffield one‐screw test. The Shapiro–Wilk test determined data normality ( p < 0.05), and nonparametric statistical tests were applied using statistical software.

Results

Descriptive statistics showed a mean linear discrepancy of 29.05 (84.80 μm) for CI and 6.95 (154.10 μm) for DI, with angular deviations of 0.06° (0.36°) for CI and 0.05° (1.40°) for DI. No statistically significant differences were found in linear ( p = 0.38) or angular ( p = 0.12) measurements between CI and DI. Framework passivity testing showed that both techniques achieved passive fit in 17 out of 20 cases (85%), with the reference scan achieving passivity in 18 (90%) cases. Distal implants, particularly in the upper jaw, exhibited greater discrepancies, but none were statistically significant.

Conclusions

No significant differences in trueness were found between CI and DI techniques. Both methods demonstrated comparable trueness and framework passivity, supporting the use of IOS as a reliable alternative to CI in edentulous jaws with multiple implants.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Jaw, EdentulousDental Impression TechniqueDental ImplantsComputer-Aided DesignHumansMaleFemaleMiddle AgedProspective StudiesAged

Stichwörter der Autor:innen

accuracyconventional impressionsdigital impressionsframework passivityimplant dentistry

Substanzen

Dental Implants

Autor:innen und Institutionen

  • Private Practice at Matanzas Clinic, Navidad, Chile.
  • Private Practice at Zenit Dental, Santiago, Chile.
  • Faculty of Dentistry, Laboratory of Periodontal Biology and Department of Pathology and Oral Medicine, University of Chile, Santiago, Chile.
  • Master Dental Technician, Division of Fixed Prosthodontics and Biomaterials, University Clinics for Dental Medicine, University of Geneva, Geneva, Switzerland.
  • Division of Fixed Prosthodontics and Biomaterials, University Clinics for Dental Medicine, University of Geneva, Geneva, Switzerland.
  • Center for Interdisciplinary Research in Health, The Catholic University of Portugal (UCP), Viseu, Portugal.

ORCID-Profile: Luís Azevedo

Förderung

  • Institute Straumann AG

Zitieren

APA

Harfagar, G., Solis, S., Hernandez, M., Fehmer, V., Sailer, I., & Azevedo, L. (2025). Trueness and Passivity of Digital and Conventional Implant Impressions in Edentulous Jaws: A Prospective Clinical Study. Clinical Implant Dentistry and Related Research. https://doi.org/10.1111/cid.70022

Vancouver

Harfagar G, Solis S, Hernandez M, Fehmer V, Sailer I, Azevedo L. Trueness and Passivity of Digital and Conventional Implant Impressions in Edentulous Jaws: A Prospective Clinical Study. Clinical Implant Dentistry and Related Research. 2025. doi:10.1111/cid.70022

BibTeX
@article{Harfagar2025,
  author = {Harfagar, Gustavo and Solis, Sebastian and Hernandez, Marcela and Fehmer, Vincent and Sailer, Irena and Azevedo, Luís},
  title = {Trueness and Passivity of Digital and Conventional Implant Impressions in Edentulous Jaws: A Prospective Clinical Study},
  journal = {Clinical Implant Dentistry and Related Research},
  year = {2025},
  doi = {10.1111/cid.70022}
}
Kostenlos starten →