Vergleichsstudie in der Zeitschrift Clinical & Experimental Ophthalmology, online erschienen am 23.06.2026. Verfasst von Jan O Weber und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Ophthalmology, Goethe-University, Frankfurt, Germany. Hauptthemen laut MEDLINE: Lenses, Intraocular, Descemet Stripping Endothelial Keratoplasty, Fuchs' Endothelial Dystrophy, Optics and Photonics. Die Arbeit nennt 24 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

To compare the predictive accuracy of ray tracing‐based intraocular lens (IOL) power calculation formulas with the Barrett Universal II (BUII) formula in eyes with Fuchs endothelial corneal dystrophy (FECD) undergoing combined cataract surgery and Descemet membrane endothelial keratoplasty (triple DMEK).

Methods

This is a retrospective, single‐centre comparative case series. Preoperative biometry was obtained using Scheimpflug tomography (Pentacam AXL) and a swept‐source optical coherence tomography–based biometer (IOLMaster 700). IOL power calculations were performed using Barrett Universal II (BUII), Olsen ray‐tracing and two OKULIX ray‐tracing approaches (paraxial and 2.5‐mm pupil). Prediction error (PE) was calculated and trueness, precision, and accuracy were evaluated using robust statistical methods (Eyetemis). A retrospective IOLup1D adjustment was additionally assessed.

Results

One hundred forty‐nine eyes of 149 patients were enrolled. BUII demonstrated the lowest trimmed mean absolute error (0.74 ± 0.65D) and median absolute error (0.60D), with 72% of eyes within ±1.00D of the target refraction. Olsen ray‐tracing showed higher errors (mean absolute error 1.06 ± 1.02D), while both OKULIX approaches performed similarly (0.91–0.92D). Only OKULIX paraxial ray‐tracing showed no significant systematic bias ( p = 0.450), whereas all other formulas demonstrated a significant hyperopic shift ( p < 0.001). After retrospective IOLup1D adjustment, BUII and Olsen achieved comparable accuracy, with Olsen demonstrating the highest trueness. Mean keratometry values measured with the IOLMaster were 0.20 ± 0.63D steeper compared to the Pentacam.

Conclusion

In triple DMEK eyes, ray tracing–based IOL power calculation did not offer better accuracy compared to BUII. BUII demonstrated the highest accuracy and precision overall. After IOLup1D adjustment, BUII and Olsen achieved comparable accuracy, with Olsen showing the greatest trueness.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Lenses, IntraocularDescemet Stripping Endothelial KeratoplastyFuchs' Endothelial DystrophyOptics and PhotonicsLens Implantation, IntraocularPhacoemulsificationHumansRetrospective StudiesBiometryAgedFemaleMaleTomography, Optical CoherenceAged, 80 and overVisual AcuityMiddle Aged

Stichwörter der Autor:innen

Barrett universal IIDescemet membrane endothelial keratoplastyFuchs endothelial corneal dystrophycataract surgeryintraocular lens power calculationray tracingtriple DMEK

Autor:innen und Institutionen

  • Department of Ophthalmology, Goethe-University, Frankfurt, Germany.
  • Department of Ophthalmology, University Hospital Munich, Ludwig-Maximilian-University, Munich, Germany.
  • Institute for Refractive and Ophthalmic Surgery (IROC), Zurich, Switzerland.
  • Department of Experimental Ophthalmology, Saarland University, Homburg, Germany.
  • Knappschaftskrankenhaus Eye Clinic, University Hospital of the Ruhr University Bochum, Bochum, Germany.

Zitieren

APA

Weber, J. O., Kaiser, K. P., Wendelstein, J. A., Dick, H. B., Kohnen, T., & Schmack, I. (2026). Ray‐Tracing‐Based Intraocular Lens Power Calculation in Combined Cataract Surgery and Descemet Membrane Endothelial Keratoplasty. Clinical & Experimental Ophthalmology, 54(6), 769-778. https://doi.org/10.1111/ceo.70135

Vancouver

Weber JO, Kaiser KP, Wendelstein JA, Dick HB, Kohnen T, Schmack I. Ray‐Tracing‐Based Intraocular Lens Power Calculation in Combined Cataract Surgery and Descemet Membrane Endothelial Keratoplasty. Clinical & Experimental Ophthalmology. 2026;54(6):769-778. doi:10.1111/ceo.70135

BibTeX
@article{Weber2026,
  author = {Weber, Jan O and Kaiser, Klemens P and Wendelstein, Jascha A and Dick, H Burkhard and Kohnen, Thomas and Schmack, Ingo},
  title = {Ray‐Tracing‐Based Intraocular Lens Power Calculation in Combined Cataract Surgery and Descemet Membrane Endothelial Keratoplasty},
  journal = {Clinical & Experimental Ophthalmology},
  year = {2026},
  volume = {54},
  number = {6},
  pages = {769--778},
  doi = {10.1111/ceo.70135}
}
Kostenlos starten →