Incorporating an Elevated Perfusion Index in the Right Hand Enhanced Screening Sensitivity for Critical Aortic Arch Obstruction in Newborn Infants
Acta Paediatrica · 2025
DOI: 10.1111/apa.17601 ↗Kurz gefasst
Fachartikel in der Zeitschrift Acta Paediatrica, 2025. Verfasst von Katarina Lannering und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Pediatric Cardiology, Queen Silvia Children's Hospital at the Sahlgrenska University Hospital, Gothenburg, Sweden. Hauptthemen laut MEDLINE: Neonatal Screening, Perfusion Index, Hand, Aorta, Thoracic. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 19 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: ALF (Agreement between Swedish government and seven regions on clinical research), Hjärt-Lungfonden.
Abstract (Originalsprache)
Aim
Newborn infants with critical aortic arch obstruction are often undiagnosed at discharge, despite screening. This study investigated if adding the perfusion index improved early detection.
Methods
We retrospectively studied 38 newborn infants with critical aortic arch obstruction, who were routinely screened in 2014–2019 by 13 Swedish hospitals using pulse oximetry and the perfusion index. They were identified through surgery records and national mortality databases. The controls were 512 healthy newborn infants from one hospital. Optimal perfusion index cut‐offs were determined using frequency distribution analysis.
Results
The groups had similar median gestational ages and birth weight. No infants with critical aortic arch obstruction were diagnosed just because of a positive perfusion index result. However, the right‐hand perfusion index was significantly higher in the cases than in controls ( p < 0.001). A perfusion index of > 3%, or positive pulse oximetry or positive physical examination, yielded 76% sensitivity and 85% specificity, with an area under the receiver operating characteristic curve of 0.81 (range 0.73–0.89, p < 0.0001). Pulse oximetry and just neonatal physical examinations had a lower sensitivity (45%, p = 0.009).
Conclusion
A high perfusion index in the right hand enhanced critical aortic arch obstruction screening and repeated measurements should be explored to minimise false positives.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Pediatric Cardiology, Queen Silvia Children's Hospital at the Sahlgrenska University Hospital, Gothenburg, Sweden.
- Department of Pediatrics, Institution of Clinical Sciences, Sahlgrenska Academy, University of Gothenburg, Gothenburg, Sweden.
ORCID-Profile: Katarina Lannering, Ingegerd Östman-Smith, Mats Mellander
Förderung
- ALF (Agreement between Swedish government and seven regions on clinical research)
- Hjärt-Lungfonden
Zitieren
APA
Lannering, K., Östman-Smith, I., & Mellander, M. (2025). Incorporating an Elevated Perfusion Index in the Right Hand Enhanced Screening Sensitivity for Critical Aortic Arch Obstruction in Newborn Infants. Acta Paediatrica. https://doi.org/10.1111/apa.17601
Vancouver
Lannering K, Östman-Smith I, Mellander M. Incorporating an Elevated Perfusion Index in the Right Hand Enhanced Screening Sensitivity for Critical Aortic Arch Obstruction in Newborn Infants. Acta Paediatrica. 2025. doi:10.1111/apa.17601
BibTeX
@article{Lannering2025,
author = {Lannering, Katarina and Östman-Smith, Ingegerd and Mellander, Mats},
title = {Incorporating an Elevated Perfusion Index in the Right Hand Enhanced Screening Sensitivity for Critical Aortic Arch Obstruction in Newborn Infants},
journal = {Acta Paediatrica},
year = {2025},
doi = {10.1111/apa.17601}
}
EUVIMED