Corneal power modelling with OCT data – Thin and thick lens paraxial models versus raytracing

Achim Langenbucher iD, Nóra Szentmáry iD, Alan Cayless, Peter Hoffmann, Jascha Wendelstein iD

Acta Ophthalmologica · 2025

DOI: 10.1111/aos.70052 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Acta Ophthalmologica, 2025. Verfasst von Achim Langenbucher und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Experimental Ophthalmology, Saarland University, Homburg (Saar), Germany. Hauptthemen laut MEDLINE: Cornea, Refraction, Ocular, Tomography, Optical Coherence, Lens, Crystalline. Die Arbeit nennt 23 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Evaluating keratometric power with Zeiss index (PKZ), paraxial thick cornea power (Gullstrand [PG]) and power referenced to the front (PFV) and back vertex plane (PBV) and raytracing power (PR), and modelling the deviation from PKZ with a multivariable linear prediction model.

Methods

A dataset of 4604 Casia2 measurements from a cataractous population was Copula expanded to N = 30 000 maintaining the individual univariate distributions and interactions. PKZ was compared with paraxial thick lens corneal power PG, PFV, PBV, and PR using measured pupil size and variations of apertures from 1 to 6 mm.

Results

On average, PKZ/PG/PFV/PBV was 43.05/42.74/42.83/43.59 D, and PR was 43.03 D with the measured pupil size. Varying pupil size from 1 (1) 6 mm increased mean PR with aperture size (42.84/42.91/43.02/43.17/43.37/43.62 D). The multivariate linear models predicting the deviation of PG‐PKZ, PFV‐PKZ, and PBV‐PKZ with corneal radii and central thickness performed well, with R 2 = 0.93 and a root mean squared prediction error of 0.01 D, whereas the equivalent model for PR‐PKZ with corneal radii and asphericities, corneal thickness, and aperture sizes performed less well, with R 2 = 0.79 and a root mean squared prediction error of 0.18 D.

Conclusion

Corneal power derived using the paraxial thick cornea model differs from keratometric power and a linear model performs well in predicting the differences. However, raytracing power differs even more from keratometry with the linear model far less effective.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

CorneaRefraction, OcularTomography, Optical CoherenceLens, CrystallineCataractHumans

Stichwörter der Autor:innen

Monte Carlo simulationcorneal powerkeratometrymultivariable linear modelparaxial thick cornea modelraytracing

Autor:innen und Institutionen

  • Department of Experimental Ophthalmology, Saarland University, Homburg (Saar), Germany.
  • Department of Ophthalmology, Semmelweis-University, Budapest, Hungary.
  • School of Physical Sciences, The Open University, Milton Keynes, UK.
  • Augen- und Laserklinik Castrop-Rauxel, Castrop-Rauxel, Germany.
  • IROC AG, Institut für Refraktive und Ophthalmo-Chirurgie, Zurich, Switzerland.
  • Department of Ophthalmology, Ludwig-Maximilian-University Munich, Munich, Germany.

ORCID-Profile: Achim Langenbucher, Nóra Szentmáry, Jascha Wendelstein

Zitieren

APA

Langenbucher, A., Szentmáry, N., Cayless, A., Hoffmann, P., & Wendelstein, J. (2025). Corneal power modelling with OCT data – Thin and thick lens paraxial models versus raytracing. Acta Ophthalmologica. https://doi.org/10.1111/aos.70052

Vancouver

Langenbucher A, Szentmáry N, Cayless A, Hoffmann P, Wendelstein J. Corneal power modelling with OCT data – Thin and thick lens paraxial models versus raytracing. Acta Ophthalmologica. 2025. doi:10.1111/aos.70052

BibTeX
@article{Langenbucher2025,
  author = {Langenbucher, Achim and Szentmáry, Nóra and Cayless, Alan and Hoffmann, Peter and Wendelstein, Jascha},
  title = {Corneal power modelling with OCT data – Thin and thick lens paraxial models versus raytracing},
  journal = {Acta Ophthalmologica},
  year = {2025},
  doi = {10.1111/aos.70052}
}
Kostenlos starten →