Influence of Refractive Index of the Crystalline Lens on Optical Biometric Measurements for Cataract Patients
Investigative Ophthalmology & Visual Science · 2026
DOI: 10.1167/iovs.67.10.61 ↗Kurz gefasst
Fachartikel in der Zeitschrift Investigative Ophthalmology & Visual Science, 2026. Verfasst von Ji C He und 7 weiteren Autor:innen. Beteiligte Einrichtung: New England College of Optometry, Boston, Massachusetts, United States. Hauptthemen laut MEDLINE: Cataract, Biometry, Lens, Crystalline, Refraction, Ocular. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: NEI NIH HHS.
Abstract (Originalsprache)
Purpose
Measurement of the axial length (AL) of the eye and the crystalline lens thickness (LT) relies on information of the refractive index of the crystalline lens for converting optical path-length (OPL) to geometric size. The purpose of this study was to measure the lens refractive index (NLens) in vivo for patients with cataract and evaluate the influence of the NLens on AL and LT measurements.
Methods
Cataract eyes of 7 patients, aged 56 to 77 years old (mean = 64.7 years), were imaged with a combined whole-eye optical coherence tomography (OCT) and Scheimpflug Imaging (SI) system before surgery. The images were processed using a custom MatLab program to trace the central light-ray from the cornea to the retina and to solve the average lens refractive index (NLens). The OPLs of other ocular segments, except the lens, were converted to geometric lengths by using the refractive indices from a model eye. The AL and LT were calculated with the measured NLens and also a constant NLens (1.41) for comparison.
Results
For the patients with cataract, mean AL and LT were 23.897 ± 0.854 and 4.309 ± 0.438 mm, respectively, when the measured NLens was used for OPL conversion, and changed to 23.918 ± 0.875 mm and 4.330 ± 0.461 mm, respectively, when the NLens was set at 1.41. The lens NLens from individual eyes varied from 1.400 to 1.434 among the patients, and caused a substantial individual variation in the AL and LT measurements, as compared with those with a constant NLens in a range from -0.070 to 0.027 mm.
Conclusions
The Scheimpflug imaging-equipped whole-eye OCT system provides a useful technique to measure AL and LT with the NLens determined for the cataract eye in vivo. The NLens varies substantially from individual eye to individual eye among the patients with cataract, and plays a critical role for deriving accurate AL and LT measurements. During cataract surgery the accurate AL and LT measurements should be used for IOL power calculation.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- New England College of Optometry, Boston, Massachusetts, United States.
- Department of Ophthalmology, Bascom Palmer Eye Institute, University of Miami Miller School of Medicine, Miami, Florida, United States.
- Duke University School of Medicine, Durham, North Carolina, United States.
- Department of Ophthalmology and Visual Sciences, Federal University of Sao Paulo - EPM/UNIFESP, Sao Paulo, SP, Brazil.
Förderung
- NEI NIH HHS
Zitieren
APA
He, J. C., Hoover, A., Gameiro, G., Moareffi, M., Ansari, Z., Tonk, R. S., Zhang, X., & Wang, J. (2026). Influence of Refractive Index of the Crystalline Lens on Optical Biometric Measurements for Cataract Patients. Investigative Ophthalmology & Visual Science. https://doi.org/10.1167/iovs.67.10.61
Vancouver
He JC, Hoover A, Gameiro G, Moareffi M, Ansari Z, Tonk RS, et al. Influence of Refractive Index of the Crystalline Lens on Optical Biometric Measurements for Cataract Patients. Investigative Ophthalmology & Visual Science. 2026. doi:10.1167/iovs.67.10.61