EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Jackknife and bootstrapping resampling techniques to evaluate the precision of lens formula constants

Achim Langenbucher iD, Jascha Wendelstein iD, Alan Cayless, Thomas Olsen, Peter Hoffmann, Nóra Szentmáry

Acta Ophthalmologica · 2025

DOI: 10.1111/aos.17522 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Acta Ophthalmologica, 2025. Verfasst von Achim Langenbucher und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Experimental Ophthalmology, Saarland University, Homburg, Saar, Germany. Hauptthemen laut MEDLINE: Lenses, Intraocular, Optics and Photonics, Refraction, Ocular, Biometry. Die Arbeit nennt 38 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

The purpose of this study was to develop a method for evaluating intraocular lens (IOL) formula constant uncertainties using two modern statistical techniques—jackknife and bootstrap resampling.

Methods

Using two datasets (dataset 1: 888 eyes treated with the aberration correcting Hoya Vivinex IOL, dataset 2: 821 eyes with the spherical Alcon SA60AT/SN60AT IOL), formula constant uncertainties for the SRK/T (Aconst), Hoffer‐Q (pACD), Holladay 1 (SF), simplified Haigis (a0) with preset a1/a2, Haigis (triplet a0/a1/a2), Castrop (triplet C/H/R) and Olsen formula (ACD) were evaluated. All input parameters were jackknife and bootstrap ( N B = 1000) resampled, and formula constants for each sample derived using nonlinear iterative optimisation techniques.

Results

In single constant formulae where the constant acts directly on the effective lens position (Hoffer‐Q, Holladay 1, simplified Haigis, Olsen), the formula constant in each case showed a standard deviation (SD) of about 0.01 with both jackknife and bootstrap sampling. The SRK/T Aconst showed a SD of about 0.018, and the Haigis and Castrop formulae with constant triplets showed large variations in the 3 constants (a0/a1/a2 about 0.036/0.005/0.002, C/H/R about 0.001/0.011/0.012). Direct formula reversion and solving for the formula constant yielded systematically larger SD values (Aconst/pACD,SF/a0/ACD = 0.586/0.395/0.403/0.324/0.304) with highly skewed distributions.

Conclusion

The distributions of formula constants with relevant benchmarks such as SD or confidence intervals can be derived with jackknife and bootstrap resampling techniques, offering potential advantages over direct formula reversion which yields skewed distributions, making central metrics such as the formula constant distribution mean unsuitable for constant optimisation.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Lenses, IntraocularOptics and PhotonicsRefraction, OcularBiometryVisual AcuityHumansProsthesis DesignLens Implantation, Intraocular

Stichwörter der Autor:innen

bootstrap resamplingconstant optimisationformula constant precisionformula constant robustnessjackknife resamplinglens power calculation

Autor:innen und Institutionen

  • Department of Experimental Ophthalmology, Saarland University, Homburg, Saar, Germany.
  • Department of Ophthalmology, Ludwig-Maximilian University Munich, Munich, Germany.
  • School of Physical Sciences, The Open University, Milton Keynes, UK.
  • Copenhagen Eye Center and Aros Private Hospital, Aarhus, Denmark.
  • Augen- und Laserklinik Castrop-Rauxel, Castrop-Rauxel, Germany.
  • Dr. Rolf M. Schwiete Center for Limbal Stem Cell and Aniridia Research, Saarland University, Homburg, Saar, Germany.

ORCID-Profile: Achim Langenbucher, Jascha Wendelstein

Zitieren

APA

Langenbucher, A., Wendelstein, J., Cayless, A., Olsen, T., Hoffmann, P., & Szentmáry, N. (2025). Jackknife and bootstrapping resampling techniques to evaluate the precision of lens formula constants. Acta Ophthalmologica. https://doi.org/10.1111/aos.17522

Vancouver

Langenbucher A, Wendelstein J, Cayless A, Olsen T, Hoffmann P, Szentmáry N. Jackknife and bootstrapping resampling techniques to evaluate the precision of lens formula constants. Acta Ophthalmologica. 2025. doi:10.1111/aos.17522

BibTeX
@article{Langenbucher2025,
  author = {Langenbucher, Achim and Wendelstein, Jascha and Cayless, Alan and Olsen, Thomas and Hoffmann, Peter and Szentmáry, Nóra},
  title = {Jackknife and bootstrapping resampling techniques to evaluate the precision of lens formula constants},
  journal = {Acta Ophthalmologica},
  year = {2025},
  doi = {10.1111/aos.17522}
}
Kostenlos starten →