Continuous Blood Gas Control Based on Active Disturbance Rejection Control During Ex Vivo Porcine Liver Perfusion
Artificial Organs · 2025
DOI: 10.1111/aor.14955 ↗At a glance
Journal article in Artificial Organs, 2025. Written by Yilong Zhao and 10 co-authors. Institution: School of Biomedical Engineering (Suzhou), Division of Life Science and Medicine, University of Science and Technology of China, Hefei, China. Major MEDLINE topics: Liver, Perfusion, Oxygen, Organ Preservation. 29 references. Funded by: Strategic Priority Research Program of the Chinese Academy of Sciences, Key Technology (Research) Project of Prevention and Control of Major Diseases and Infectious Diseases of Suzhou, China.
Abstract (original language)
Background
Membrane oxygenators facilitate extracorporeal gas exchange, necessitating the monitoring of blood gas. Recent advances in normothermic machine perfusion (NMP) for ex vivo liver offer solutions to the shortage of donor liver. However, maintaining physiological blood gas levels during prolonged NMP is complex and costly.
Methods
We introduce a noninvasive and economical approach for regulating the blood gas during NMP of ex vivo porcine livers. By monitoring gas fractions at the outlet of oxygenator, real‐time adjustments of blood gas can be made without the online blood gas analyzer. The method involves constructing multivariate linear regression (MLR) models, aligning target setpoints of gas, and employing active disturbance rejection control (ADRC) to achieve closed‐loop regulation.
Results
Ex vivo porcine liver perfusion experiments demonstrated the effectiveness of the method, maintaining blood gas within physiological levels over 24 h (oxygen partial pressure: 150.36 ± 3.33 mmHg, carbon dioxide partial pressure: 41.34 ± 0.91 mmHg).
Conclusion
ADRC‐based continuous regulation of gas fraction at the outlet of oxygenator is a feasible and effective approach for managing blood gas during ex vivo porcine liver perfusion.
Subjects
MeSH terms (MEDLINE)
Author keywords
Substances
Authors and institutions
- School of Biomedical Engineering (Suzhou), Division of Life Science and Medicine, University of Science and Technology of China, Hefei, China.
- Suzhou Institute of Biomedical Engineering and Technology, Chinese Academy of Sciences, Suzhou, China.
- State Key Laboratory of Cell Biology, Center for Excellence in Molecular Cell Science, Shanghai Institute of Biochemistry and Cell Biology, Chinese Academy of Sciences, University of Chinese Academy of Sciences, Shangha…
- Hainan Academy of Medical Sciences, Hainan Medical University, Haikou, Hainan, China.
ORCID profiles: Yilong Zhao, Gangyin Luo
Funding
- Strategic Priority Research Program of the Chinese Academy of Sciences
- Key Technology (Research) Project of Prevention and Control of Major Diseases and Infectious Diseases of Suzhou, China
Cite
APA
Zhao, Y., Lyu, X., Sun, Z., Zhang, X., Cen, J., Yang, T., Xu, X., Xing, W., Zhao, S., Wang, B., & Luo, G. (2025). Continuous Blood Gas Control Based on Active Disturbance Rejection Control During Ex Vivo Porcine Liver Perfusion. Artificial Organs. https://doi.org/10.1111/aor.14955
Vancouver
Zhao Y, Lyu X, Sun Z, Zhang X, Cen J, Yang T, et al. Continuous Blood Gas Control Based on Active Disturbance Rejection Control During Ex Vivo Porcine Liver Perfusion. Artificial Organs. 2025. doi:10.1111/aor.14955
BibTeX
@article{Zhao2025,
author = {Zhao, Yilong and Lyu, Xin and Sun, Zhen and Zhang, Xiaoliang and Cen, Jin and Yang, Tianhang and Xu, Xiaoliang and Xing, Wenhui and Zhao, Sihan and Wang, Bidou and Luo, Gangyin},
title = {Continuous Blood Gas Control Based on Active Disturbance Rejection Control During Ex Vivo Porcine Liver Perfusion},
journal = {Artificial Organs},
year = {2025},
doi = {10.1111/aor.14955}
}
EUVIMED