EUVIMEDOpdag EUVIMED
Evaluationsstudie

Performance evaluation of an automatically controlled transcutaneous energy transfer system powering left ventricular assist device

Sarath S Nair iD, D S Nagesh, Sachin J Shenoy, S Harikrishnan

Artificial Organs · 2025

DOI: 10.1111/aor.14892 ↗

At a glance

Evaluationsstudie in Artificial Organs, 2025. Written by Sarath S Nair and 3 co-authors. Institution: Department of Medical Devices Engineering, Biomedical Technology Wing, Sree Chitra Tirunal Institute for Medical Sciences and Technology (SCTIMST), G…. Major MEDLINE topics: Heart-Assist Devices, Energy Transfer, Electric Power Supplies. 25 references. Funded by: INAE-SERB(DST) Abdul Kalam Technology Innovation National Fellowship.

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Background

The survival rate of LVAD recipients may be improved by eliminating the infection failure mode at the percutaneous lead entry site. Wireless powering through a Transcutaneous Energy Transfer System (TETS) is a promising solution for achieving this. However, automatic controls need to be employed to compensate for the variations in efficiency and power transfer due to changes in load and coil‐to‐coil gaps.

Methods

This article discusses the results of in vitro and animal models performance evaluation of an automatically controlled TETS for powering a Left Ventricular Assist Device.

Results

It was found that power up to 20 W could be transferred at an axial gap of 30 mm. A 5% variation in speed and a 2% change in flow rate are observed with the change in tissue thickness, in porcine carcass model.

Conclusions

The LVAD could be operated at the required RPM irrespective of the change in axial positions.

Subjects

MeSH terms (MEDLINE)

Heart-Assist DevicesEnergy TransferElectric Power SuppliesAnimalsSwineWireless TechnologyProsthesis Design

Author keywords

Total artificial heartleft ventricular assist devicetranscutaneous energy transfer wireless energy transfer

Authors and institutions

  • Department of Medical Devices Engineering, Biomedical Technology Wing, Sree Chitra Tirunal Institute for Medical Sciences and Technology (SCTIMST), Government of India, Thiruvananthapuram, Kerala, India.
  • Department of Applied Biology, Biomedical Technology Wing, Sree Chitra Tirunal Institute for Medical Sciences and Technology (SCTIMST), Government of India, Thiruvananthapuram, Kerala, India.
  • Department of Cardiology, Sree Chitra Tirunal Institute for Medical Sciences and Technology (SCTIMST), Government of India, Thiruvananthapuram, Kerala, India.

ORCID profiles: Sarath S Nair

Funding

  • INAE-SERB(DST) Abdul Kalam Technology Innovation National Fellowship

Cite

APA

Nair, S. S., Nagesh, D. S., Shenoy, S. J., & Harikrishnan, S. (2025). Performance evaluation of an automatically controlled transcutaneous energy transfer system powering left ventricular assist device. Artificial Organs. https://doi.org/10.1111/aor.14892

Vancouver

Nair SS, Nagesh DS, Shenoy SJ, Harikrishnan S. Performance evaluation of an automatically controlled transcutaneous energy transfer system powering left ventricular assist device. Artificial Organs. 2025. doi:10.1111/aor.14892

BibTeX
@article{Nair2025,
  author = {Nair, Sarath S and Nagesh, D S and Shenoy, Sachin J and Harikrishnan, S},
  title = {Performance evaluation of an automatically controlled transcutaneous energy transfer system powering left ventricular assist device},
  journal = {Artificial Organs},
  year = {2025},
  doi = {10.1111/aor.14892}
}
Start for free →