Performance evaluation of an automatically controlled transcutaneous energy transfer system powering left ventricular assist device
Artificial Organs · 2025
DOI: 10.1111/aor.14892 ↗At a glance
Evaluationsstudie in Artificial Organs, 2025. Written by Sarath S Nair and 3 co-authors. Institution: Department of Medical Devices Engineering, Biomedical Technology Wing, Sree Chitra Tirunal Institute for Medical Sciences and Technology (SCTIMST), G…. Major MEDLINE topics: Heart-Assist Devices, Energy Transfer, Electric Power Supplies. 25 references. Funded by: INAE-SERB(DST) Abdul Kalam Technology Innovation National Fellowship.
Abstract (original language)
Background
The survival rate of LVAD recipients may be improved by eliminating the infection failure mode at the percutaneous lead entry site. Wireless powering through a Transcutaneous Energy Transfer System (TETS) is a promising solution for achieving this. However, automatic controls need to be employed to compensate for the variations in efficiency and power transfer due to changes in load and coil‐to‐coil gaps.
Methods
This article discusses the results of in vitro and animal models performance evaluation of an automatically controlled TETS for powering a Left Ventricular Assist Device.
Results
It was found that power up to 20 W could be transferred at an axial gap of 30 mm. A 5% variation in speed and a 2% change in flow rate are observed with the change in tissue thickness, in porcine carcass model.
Conclusions
The LVAD could be operated at the required RPM irrespective of the change in axial positions.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Department of Medical Devices Engineering, Biomedical Technology Wing, Sree Chitra Tirunal Institute for Medical Sciences and Technology (SCTIMST), Government of India, Thiruvananthapuram, Kerala, India.
- Department of Applied Biology, Biomedical Technology Wing, Sree Chitra Tirunal Institute for Medical Sciences and Technology (SCTIMST), Government of India, Thiruvananthapuram, Kerala, India.
- Department of Cardiology, Sree Chitra Tirunal Institute for Medical Sciences and Technology (SCTIMST), Government of India, Thiruvananthapuram, Kerala, India.
ORCID profiles: Sarath S Nair
Funding
- INAE-SERB(DST) Abdul Kalam Technology Innovation National Fellowship
Cite
APA
Nair, S. S., Nagesh, D. S., Shenoy, S. J., & Harikrishnan, S. (2025). Performance evaluation of an automatically controlled transcutaneous energy transfer system powering left ventricular assist device. Artificial Organs. https://doi.org/10.1111/aor.14892
Vancouver
Nair SS, Nagesh DS, Shenoy SJ, Harikrishnan S. Performance evaluation of an automatically controlled transcutaneous energy transfer system powering left ventricular assist device. Artificial Organs. 2025. doi:10.1111/aor.14892
BibTeX
@article{Nair2025,
author = {Nair, Sarath S and Nagesh, D S and Shenoy, Sachin J and Harikrishnan, S},
title = {Performance evaluation of an automatically controlled transcutaneous energy transfer system powering left ventricular assist device},
journal = {Artificial Organs},
year = {2025},
doi = {10.1111/aor.14892}
}
EUVIMED