Reduced Heating Wireless Energy Transmission System for Powering Implanted Circulatory Assist Devices: Benchtop and In-Vivo Studies
Sensors · 2025
DOI: 10.3390/s25051311 ↗Kurz gefasst
Fachartikel in der Zeitschrift Sensors, 2025. Verfasst von Mohammad L Karim und 7 weiteren Autor:innen. Beteiligte Einrichtung: Nanotechnology & Integrated BioEngineering Centre (NIBEC), School of Engineering, Ulster University, Belfast BT15 1AP, UK. Hauptthemen laut MEDLINE: Heart-Assist Devices, Wireless Technology, Electric Power Supplies. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 28 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Innovate UK, UK Research and Innovation (UKRI).
Abstract (Originalsprache)
This study aimed to develop a novel Transdermal Energy Transmission System (TETS) device that addresses the driveline complications faced by patients with advanced heart failure (HF). Our TETS device utilizes a two-channel configuration with a very-low duty cycle and a pulsed RF power transmission technique, along with elliptically shaped flexible coil inductive coupling elements. We integrated a battery charging controller module into the TETS, enabling it to recharge an implanted Lithium-Ion (Li-Ion) battery that powers low-power-rated Circulatory Assist Devices, or left ventricular assist devices (LVADs). Benchtop measurements demonstrated that the TETS delivered energy from the implanted coils to the battery charging module, at a charging rate of up to 2900 J/h, presented an average temperature increase (ΔT) of 3 °C. We conducted in vivo measurements using four porcine models followed by histopathological analysis of the skin tissue in the implanted coils areas. The thermal profile analysis from the in vivo measurements and the calculated charging rates from the current and voltage waveforms, in porcine models, indicated that the charging rate and temperature varied for each model. The maximum energy charging rate observed was 2200 J/h, with an average ΔT of 3 °C. The exposed skin tissue histopathological analysis results showed no evidence of tissue thermal damage in the in vivo measurements. These results demonstrate the feasibility of our developed TETS device for wireless driving implanted low-power-rated LVADs and Li-Ion charging.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Nanotechnology & Integrated BioEngineering Centre (NIBEC), School of Engineering, Ulster University, Belfast BT15 1AP, UK.
- Galvani TECH Ltd., School of Engineering, Ulster University, Belfast BT15 1AP, UK.
- Paul Crawford Veterinary Services, Larne BT40 3RW, UK.
- Cardiovascular Research Unit, Craigavon Area Hospital, Portadown BT63 5QQ, UK.
ORCID-Profile: Mohammad L Karim, James McLaughlin, Paul Crawford, Omar J Escalona
Förderung
- Innovate UK
- UK Research and Innovation (UKRI)
Zitieren
APA
Karim, M. L., Grimes, R., Larkin, H., Bosnjak, A. M., McLaughlin, J., Crawford, P., McEneaney, D., & Escalona, O. J. (2025). Reduced Heating Wireless Energy Transmission System for Powering Implanted Circulatory Assist Devices: Benchtop and In-Vivo Studies. Sensors. https://doi.org/10.3390/s25051311
Vancouver
Karim ML, Grimes R, Larkin H, Bosnjak AM, McLaughlin J, Crawford P, et al. Reduced Heating Wireless Energy Transmission System for Powering Implanted Circulatory Assist Devices: Benchtop and In-Vivo Studies. Sensors. 2025. doi:10.3390/s25051311
BibTeX
@article{Karim2025,
author = {Karim, Mohammad L and Grimes, Rachel and Larkin, Harry and Bosnjak, Antonio M and McLaughlin, James and Crawford, Paul and McEneaney, David and Escalona, Omar J},
title = {Reduced Heating Wireless Energy Transmission System for Powering Implanted Circulatory Assist Devices: Benchtop and In-Vivo Studies},
journal = {Sensors},
year = {2025},
doi = {10.3390/s25051311}
}