Comparison of hemodialysis urea clearance using spent dialysate and Kt/Vurea equations
Artificial Organs · 2025
DOI: 10.1111/aor.14848 ↗At a glance
Vergleichsstudie in Artificial Organs, 2025. Written by Priyanka Khatri and 1 co-authors. Institution: Fast and Chronic Programmes, Alexandra Hospital, Singapore, Singapore. Major MEDLINE topics: Urea, Renal Dialysis, Dialysis Solutions, Hemodiafiltration. Cited 2 times so far (Crossref); 27 references. The full text is freely available (open access, CC BY 4.0).
Abstract (original language)
Introduction
Dialysis adequacy is traditionally calculated from pre‐ and post‐hemodialysis session serum urea concentrations and expressed as the urea reduction ratio, or Kt/Vurea. However, with increasing hemodiafiltration usage, we wished to determine whether there were any differences between standard Kt/Vurea equations and directly measured spent dialysate urea clearance.
Methods
Urea clearance was measured from collected effluent dialysate and compared with various other methods of Kt/Vurea calculation, including change in total body urea from measuring pre‐ and post‐total body water with bioimpedance and the Watson equation, by standard Kt/V equations, and online clearance measurements using effective ionic dialysance (OLC).
Results
We compared urea clearance in 41 patients, 56.1% male, mean age 69.3 ± 12.6 years with 87.8% treated by hemodiafiltration. Reduction in total body urea was greater when estimating changes in total body urea, compared to measured dialysate losses of 58.4% (48.5–67.6) vs 71.6% (62.1–78), p < 0.01. Sessional urea clearance (Kt/Vurea) was greater using the online Solute‐Solver program compared to OLC, median 1.45(1.13–1.75) vs 1.2 (0.93–1.4), and 2nd generation Kt/V equations 1.3 (1.02–1.66), p < 0.01, but not different from estimated total body urea clearance 1.36 (1.15–1.73) and dialysate clearance 1.36 (1.07–1.76). The mean bias compared to the Solute‐Solver program was greatest with OLC (−0.25), compared to second‐generation equations (−0.02), estimated total body clearance (−0.02) and measured dialysate clearance (−0.01).
Conclusion
This study demonstrated that the result from equations estimating urea clearance indirectly from pre‐ and postblood samples from hemo‐ and hemodiafiltration treatments was highly correlated with direct measurements of dialysate urea clearance.
Subjects
MeSH terms (MEDLINE)
Author keywords
Substances
Authors and institutions
- Fast and Chronic Programmes, Alexandra Hospital, Singapore, Singapore.
- Yong Loo Lin School of Medicine, National University of Singapore, Singapore, Singapore.
- UCL Department of Renal Medicine, Royal Free Hospital, London, UK.
Cite
APA
Khatri, P., & Davenport, A. (2025). Comparison of hemodialysis urea clearance using spent dialysate and Kt/Vurea equations. Artificial Organs. https://doi.org/10.1111/aor.14848
Vancouver
Khatri P, Davenport A. Comparison of hemodialysis urea clearance using spent dialysate and Kt/Vurea equations. Artificial Organs. 2025. doi:10.1111/aor.14848
BibTeX
@article{Khatri2025,
author = {Khatri, Priyanka and Davenport, Andrew},
title = {Comparison of hemodialysis urea clearance using spent dialysate and Kt/Vurea equations},
journal = {Artificial Organs},
year = {2025},
doi = {10.1111/aor.14848}
}
EUVIMED