Mid-Range Dialysate Flow and Adequacy of Dialysis
Blood Purification · 2025
DOI: 10.1159/000545703 ↗Kurz gefasst
Fachartikel in der Zeitschrift Blood Purification, 2025. Verfasst von Ami M Patel und 3 weiteren Autor:innen. Beteiligte Einrichtung: Division of Nephrology, University of Maryland School of Medicine, Baltimore, Maryland, USA. Hauptthemen laut MEDLINE: Renal Dialysis, Acute Kidney Injury, Kidney Failure, Chronic, Dialysis Solutions. Die Arbeit nennt 23 Literaturangaben.
Abstract (Originalsprache)
Introduction
The primary objective was to investigate the efficiency of a fully integrated, hemodialysis machine that utilizes mid-range dialysate flow rate of 300 mL/min in the acute hospital setting.
Methods
We conducted a retrospective, single-center pilot study. All veterans with end-stage kidney disease or dialysis-dependent acute kidney injury, who completed ≥3 h of hemodialysis and had pre-dialysis blood urea nitrogen (BUN) >20 mg/dL with available post-dialysis BUN from February 2023 to June 2023 were included. Urea reduction ratio (URR) and single-pool Kt/V were compared between groups with varying dialyzer sizes and duration of treatment.
Results
Fifty-nine dialysis sessions were included for analysis. Mean Kt/V and URR for the full cohort were 1.26 ± 0.04 and 60.8 ± 7.3%, respectively. The percentage of sessions achieving minimum adequate clearance per Kt/V and URR criteria were 50.6% and 32.5%, respectively. In the subgroup analysis of dialysis sessions of different filter size (Optiflux™ F160NR or F180NR dialyzers) and varying duration (3, 3.5, 4 h), the percentage of sessions achieving minimum adequate Kt/V in the F160/3.5 h, F180/3.5 h and F180/4 h groups were 44%, 50%, and 58.8%, respectively.
Conclusion
The restriction of QD to 300 mL/min may compromise adequacy of dialysis in acute care setting. This may be mitigated by prolonging session times, albeit at the expense of higher resource utilization, and improving the QD:QB ratio. Further studies in a larger hospital cohort are necessary to elucidate these issues.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Division of Nephrology, University of Maryland School of Medicine, Baltimore, Maryland, USA.
- VA Maryland Health Care System, Baltimore, Maryland, USA.
Zitieren
APA
Patel, A. M., Atefi, N., Chahal, J., & Ejaz, A. A. (2025). Mid-Range Dialysate Flow and Adequacy of Dialysis. Blood Purification. https://doi.org/10.1159/000545703
Vancouver
Patel AM, Atefi N, Chahal J, Ejaz AA. Mid-Range Dialysate Flow and Adequacy of Dialysis. Blood Purification. 2025. doi:10.1159/000545703
BibTeX
@article{Patel2025,
author = {Patel, Ami M and Atefi, Nazli and Chahal, Jagman and Ejaz, Abutaleb Ahsan},
title = {Mid-Range Dialysate Flow and Adequacy of Dialysis},
journal = {Blood Purification},
year = {2025},
doi = {10.1159/000545703}
}