Mid-Range Dialysate Flow and Adequacy of Dialysis

Ami M Patel, Nazli Atefi, Jagman Chahal, Abutaleb Ahsan Ejaz

Blood Purification · 2025

DOI: 10.1159/000545703 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Blood Purification, 2025. Verfasst von Ami M Patel und 3 weiteren Autor:innen. Beteiligte Einrichtung: Division of Nephrology, University of Maryland School of Medicine, Baltimore, Maryland, USA. Hauptthemen laut MEDLINE: Renal Dialysis, Acute Kidney Injury, Kidney Failure, Chronic, Dialysis Solutions. Die Arbeit nennt 23 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

The primary objective was to investigate the efficiency of a fully integrated, hemodialysis machine that utilizes mid-range dialysate flow rate of 300 mL/min in the acute hospital setting.

Methods

We conducted a retrospective, single-center pilot study. All veterans with end-stage kidney disease or dialysis-dependent acute kidney injury, who completed ≥3 h of hemodialysis and had pre-dialysis blood urea nitrogen (BUN) >20 mg/dL with available post-dialysis BUN from February 2023 to June 2023 were included. Urea reduction ratio (URR) and single-pool Kt/V were compared between groups with varying dialyzer sizes and duration of treatment.

Results

Fifty-nine dialysis sessions were included for analysis. Mean Kt/V and URR for the full cohort were 1.26 ± 0.04 and 60.8 ± 7.3%, respectively. The percentage of sessions achieving minimum adequate clearance per Kt/V and URR criteria were 50.6% and 32.5%, respectively. In the subgroup analysis of dialysis sessions of different filter size (Optiflux™ F160NR or F180NR dialyzers) and varying duration (3, 3.5, 4 h), the percentage of sessions achieving minimum adequate Kt/V in the F160/3.5 h, F180/3.5 h and F180/4 h groups were 44%, 50%, and 58.8%, respectively.

Conclusion

The restriction of QD to 300 mL/min may compromise adequacy of dialysis in acute care setting. This may be mitigated by prolonging session times, albeit at the expense of higher resource utilization, and improving the QD:QB ratio. Further studies in a larger hospital cohort are necessary to elucidate these issues.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Renal DialysisAcute Kidney InjuryKidney Failure, ChronicDialysis SolutionsHumansRetrospective StudiesMaleFemaleMiddle AgedAgedPilot ProjectsBlood Urea NitrogenUrea

Stichwörter der Autor:innen

Hemodialysis adequacyKt/VMid-range dialysate flow rateUrea reduction ratio

Substanzen

Dialysis SolutionsUrea

Autor:innen und Institutionen

  • Division of Nephrology, University of Maryland School of Medicine, Baltimore, Maryland, USA.
  • VA Maryland Health Care System, Baltimore, Maryland, USA.

Zitieren

APA

Patel, A. M., Atefi, N., Chahal, J., & Ejaz, A. A. (2025). Mid-Range Dialysate Flow and Adequacy of Dialysis. Blood Purification. https://doi.org/10.1159/000545703

Vancouver

Patel AM, Atefi N, Chahal J, Ejaz AA. Mid-Range Dialysate Flow and Adequacy of Dialysis. Blood Purification. 2025. doi:10.1159/000545703

BibTeX
@article{Patel2025,
  author = {Patel, Ami M and Atefi, Nazli and Chahal, Jagman and Ejaz, Abutaleb Ahsan},
  title = {Mid-Range Dialysate Flow and Adequacy of Dialysis},
  journal = {Blood Purification},
  year = {2025},
  doi = {10.1159/000545703}
}
Kostenlos starten →