Optimization design of semi‐open impeller based on thrombogenicity in a blood pump

Xuemin Liu iD, Jiejie Shao, Peng Wang, Honghua Zhao, Luning Liu, Qing Han iD

Artificial Organs · 2024

DOI: 10.1111/aor.14813 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Artificial Organs, 2024. Verfasst von Xuemin Liu und 5 weiteren Autor:innen. Beteiligte Einrichtung: School of Mechanical Engineering, University of Jinan, Jinan, China. Hauptthemen laut MEDLINE: Thrombosis, Heart-Assist Devices, Blood Platelets. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 40 Literaturangaben. Gefördert von: Natural Science Foundation of Shandong Province.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Blood clots are composed of aggregated fibrin and platelets, and thrombosis is the body's natural response to repairing injured blood vessels or stopping bleeding. However, when this process is activated abnormally, such as in a mechanical blood pump, it can lead to excessive thrombus formation. Therefore, how to avoid or reduce the probability of thrombus formation is an important indicator of the stable operation of a blood pump.

Methods

In this paper, Lagrangian particle tracking trajectories are simulated to study platelet transport in a blood pump. The design of the thrombus blood pump was optimized using an orthogonal design method based on three factors: inlet angle, outlet angle, and blade number. The effect of blood pump pressure, rotational speed, impeller outlet angle, inlet angle, and number of blades on thrombus formation was analysed using Fluent software. The thrombogenic potential was derived by analyzing the trajectory and flow parameters of platelet particles in the blood pump, as well as the statistical parameters of residence time and stress accumulation thrombus in the platelet pump.

Results

When the impeller inlet angle is 30°, the outlet angle is 20°, and the number of blades is 6, the probability of thrombus formation is minimized in the orthogonal design method, aligning with the requirements for blood pump performance.

Conclusions

These design parameters serve as a numerical guideline for optimizing the geometry of the semi‐open impeller in blood pumps and provide a theoretical foundation for subsequent in vitro experiments.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ThrombosisHeart-Assist DevicesBlood PlateletsHumansModels, CardiovascularComputer SimulationEquipment Design

Stichwörter der Autor:innen

blood pumpcomputational fluid dynamicsoptimization designsemi‐open impellerthrombogenicity

Autor:innen und Institutionen

  • School of Mechanical Engineering, University of Jinan, Jinan, China.
  • Ningbo Zhongjie Laitong Technology Co., Ltd., Ningbo, China.

ORCID-Profile: Xuemin Liu, Qing Han

Förderung

  • Natural Science Foundation of Shandong Province

Zitieren

APA

Liu, X., Shao, J., Wang, P., Zhao, H., Liu, L., & Han, Q. (2024). Optimization design of semi‐open impeller based on thrombogenicity in a blood pump. Artificial Organs. https://doi.org/10.1111/aor.14813

Vancouver

Liu X, Shao J, Wang P, Zhao H, Liu L, Han Q. Optimization design of semi‐open impeller based on thrombogenicity in a blood pump. Artificial Organs. 2024. doi:10.1111/aor.14813

BibTeX
@article{Liu2024,
  author = {Liu, Xuemin and Shao, Jiejie and Wang, Peng and Zhao, Honghua and Liu, Luning and Han, Qing},
  title = {Optimization design of semi‐open impeller based on thrombogenicity in a blood pump},
  journal = {Artificial Organs},
  year = {2024},
  doi = {10.1111/aor.14813}
}
Kostenlos starten →