Longitudinal cardiac dimensions in patients undergoing LVAD implantation
Artificial Organs · 2024
DOI: 10.1111/aor.14728 ↗At a glance
Journal article in Artificial Organs, 2024. Written by Florian Meissner and 7 co-authors. Institution: Department of Cardiovascular Surgery, Heart Center Freiburg, Faculty of Medicine, University of Freiburg, Freiburg, Germany. Major MEDLINE topics: Thoracic Surgical Procedures, Heart-Assist Devices, Heart Failure. Cited 3 times so far (Crossref); 19 references. The full text is freely available (open access, CC BY-NC 4.0). Funded by: German Federal Ministry of Education and Research, Bundesministerium für Bildung und Forschung.
Abstract (original language)
Background
In conventional left ventricular assist devices (LVAD), a separate outflow graft is sutured to the ascending aorta. Novel device designs may include a transventricular outflow cannula crossing the aortic valve (AV). While transversal ventricular dimensions are well investigated in patients with severe heart failure, little is known about the longitudinal dimensions. These dimensions are, however, particularly critical for the design and development of mechanical circulatory support (MCS) devices with transaortic outflow cannula.
Methods
In an explorative retrospective cohort study at the University Medical Center Freiburg, Germany, the longitudinal cardiac dimensions of patients undergoing computed tomography angiography (CTA) before and, if available, after LVAD implantation were analyzed. Among others, the following dimensions were assessed: (a) apex to AV, (b) apex to mitral valve, (c) AV to sinotubular junction (STJ), (d) apex to STJ, (e) apex to brachiocephalic artery (BCA), and (f) AV to BCA.
Results
In total, 44 LVAD patients (36 male, age 55.8 years, height 1.75 m) were included. The longitudinal cardiac dimensions were (a) 114.5 ± 12.1 mm, (b) 108.0 ± 12.4 mm, (c) 20.9 ± 2.9, (d) 135.4 ± 13.4 mm, (e) 206.0 ± 18.3, and (f) 91.5 ± 9.8 mm. Postoperatively, (a) and (b) decreased by 31.5% and 39.5%, respectively ( N = 14).
Conclusions
Longitudinal cardiac dimensions may be reduced by up to 40% after LVAD implantation. A better knowledge of these dimensions and their postoperative alterations in LVAD patients may improve surgical planning and help to design MCS devices with transventricular outflow cannula.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Department of Cardiovascular Surgery, Heart Center Freiburg, Faculty of Medicine, University of Freiburg, Freiburg, Germany.
- Department of Diagnostic and Interventional Radiology, Faculty of Medicine, University of Freiburg, Freiburg, Germany.
- Institute of Medical Biometry and Statistics, Faculty of Medicine, University of Freiburg, Freiburg, Germany.
ORCID profiles: Florian Meissner, David Schibilsky
Funding
- German Federal Ministry of Education and Research
- Bundesministerium für Bildung und Forschung
Cite
APA
Meissner, F., Szvetics, S., Galbas, M. C., Russe, M., Schibilsky, D., Kaier, K., Czerny, M., & Bothe, W. (2024). Longitudinal cardiac dimensions in patients undergoing LVAD implantation. Artificial Organs. https://doi.org/10.1111/aor.14728
Vancouver
Meissner F, Szvetics S, Galbas MC, Russe M, Schibilsky D, Kaier K, et al. Longitudinal cardiac dimensions in patients undergoing LVAD implantation. Artificial Organs. 2024. doi:10.1111/aor.14728
BibTeX
@article{Meissner2024,
author = {Meissner, Florian and Szvetics, Sophie and Galbas, Michelle Costa and Russe, Maximilian and Schibilsky, David and Kaier, Klaus and Czerny, Martin and Bothe, Wolfgang},
title = {Longitudinal cardiac dimensions in patients undergoing LVAD implantation},
journal = {Artificial Organs},
year = {2024},
doi = {10.1111/aor.14728}
}
EUVIMED