Experimental validation of the power law hemolysis model using a Couette shearing device
Artificial Organs · 2024-05 · 48 (5) · 495-503
DOI: 10.1111/aor.14702 ↗Kurz gefasst
Fachartikel in der Zeitschrift Artificial Organs, online erschienen am 26.12.2023. Verfasst von Vera Charlotte Lommel und 3 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Computer-assisted Cardiovascular Medicine, Biofluid Mechanics Laboratory, Deutsches Herzzentrum der Charité (DHZC), Berlin, Germany. Hauptthemen laut MEDLINE: Heart-Assist Devices. Die Arbeit wurde bisher 13-mal zitiert (Crossref) und nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: European Regional Development Fund.
Abstract (Originalsprache)
Background
The study of blood trauma, such as hemolysis in blood‐carrying devices, is crucial due to the high incidence of adverse events like alteration of blood function, bleeding, and multi‐organ failure. The extent of flow‐induced hemolysis, predominantly influenced by stress duration and intensity, is described by established model parameters based on the power law approach. In recent years, various parameters were determined using different Couette shearing devices and donor species. However, they have not been validated due to limited experimental data.
Methods
This study provides hemolysis measurements in a Couette shearing device and evaluates the suitability of different power law parameters. The revised Couette shearing device generates well‐defined dynamic stress loads that are repeatedly applied to blood samples at a defined temperature. Human blood samples with an adjusted hematocrit of 30%, were tested with varying repetitions (20 to 80 times). The half‐sinusoidal stress loads had amplitudes of 73 to 140 Pa and exposure times of 24 msec per repetition. The parameters of five common power law hemolysis approaches were then compared with the experimental data.
Results
The prediction with the power law model parameters C = 3.458 × 10 −6 , α = 0.2777 and β = 2.0639 showed a good agreement with the experimental results.
Conclusion
The effect of multiple short‐time stresses on hemolysis was investigated to validate the power law hemolysis model with the Couette shearing device of this study.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Institute of Computer-assisted Cardiovascular Medicine, Biofluid Mechanics Laboratory, Deutsches Herzzentrum der Charité (DHZC), Berlin, Germany.
- Charité-Universitätsmedizin Berlin, Corporate Member of Freie Universität Berlin and Humboldt-Universität zu Berlin, Berlin, Germany.
ORCID-Profile: Vera Charlotte Lommel, Leonid Goubergrits, Ulrich Kertzscher, Michael Lommel
Förderung
- European Regional Development Fund
Zitieren
APA
Lommel, V. C., Goubergrits, L., Kertzscher, U., & Lommel, M. (2024). Experimental validation of the power law hemolysis model using a Couette shearing device. Artificial Organs, 48(5), 495-503. https://doi.org/10.1111/aor.14702
Vancouver
Lommel VC, Goubergrits L, Kertzscher U, Lommel M. Experimental validation of the power law hemolysis model using a Couette shearing device. Artificial Organs. 2024;48(5):495-503. doi:10.1111/aor.14702
BibTeX
@article{Lommel2024,
author = {Lommel, Vera Charlotte and Goubergrits, Leonid and Kertzscher, Ulrich and Lommel, Michael},
title = {Experimental validation of the power law hemolysis model using a Couette shearing device},
journal = {Artificial Organs},
year = {2024},
volume = {48},
number = {5},
pages = {495--503},
doi = {10.1111/aor.14702}
}
EUVIMED