Preservation of pulsatility with universal ventricular assist device: In vitro assessment for biventricular support

Taiyo Kuroda, Chihiro Miyagi iD, Anthony R Polakowski iD, Christine R Flick, Barry D Kuban, Kiyotaka Fukamachi iD, Jamshid H Karimov iD

Artificial Organs · 2024

DOI: 10.1111/aor.14656 ↗

At a glance

Journal article in Artificial Organs, 2024. Written by Taiyo Kuroda and 6 co-authors. Institution: Department of Biomedical Engineering, Lerner Research Institute, Cleveland Clinic, Cleveland, Ohio, USA. Major MEDLINE topics: Heart-Assist Devices, Heart Failure. 20 references.

AstrattSubjectsDettalji biblijografiċiCiteRelated studies

Abstract (original language)

Background

The objective of this study was to assess the pulsatility preservation capability of the universal ventricular assist device (UVAD) when used as a biventricular assist device (BVAD). This evaluation was conducted through an in vitro experiment, utilizing a pulsatile biventricular circulatory mock loop.

Methods

Two UVAD pumps were tested in a dual setup (BVAD) in the circulatory model with the simulated conditions of left heart failure (HF), right HF, and moderate/severe biventricular HF (BHF). The total flow, aortic pulse pressure, the pulse augmentation factor (PAF), the energy‐equivalent pressure (EEP), and the surplus hemodynamic energy (SHE) were observed at various pump speeds to evaluate the pulsatility.

Results

The aortic pulse pressure increased from the baseline (without pump) in all simulated hemodynamic conditions. The PAF ranged from 17%–35% in healthy, left HF, right HF, and mild BHF conditions, with the highest PAF of 90% being observed in the severe BHF condition. The EEP correlated with LVAD flow in all groups ( R 2 = 0.87–0.97) and increased from the baseline in all cases. The SHE peaked at approximately 5–6 L/min of LVAD support and was likely to decrease at higher LVAD pump flow. The largest decrease in SHE from the baseline, 53%, was observed in the mild BHF conditions with the highest LVAD and RVAD support.

Conclusions

The UVAD successfully demonstrated the ability to preserve pulsatility in vitro, and to optimize the cardiac output, as an isolated circulatory support device option (RVAD or LVAD) and when used for BVAD support.

Subjects

MeSH terms (MEDLINE)

Heart-Assist DevicesHeart FailureHumansChinaEthnicityHemodynamicsCardiac Output

Author keywords

biventricular heart failureblood pumpscardiovascular simulation modelmechanical circulatory supportuniversal deviceventricular assist device

Authors and institutions

  • Department of Biomedical Engineering, Lerner Research Institute, Cleveland Clinic, Cleveland, Ohio, USA.
  • Department of Cardiovascular Medicine, Kaufman Center for Heart Failure Treatment and Recovery, Section of Heart Failure and Cardiac Transplant Medicine, Cleveland, Ohio, USA.
  • Cleveland Clinic Lerner College of Medicine of Case Western Reserve University, Cleveland, Ohio, USA.

ORCID profiles: Chihiro Miyagi, Anthony R Polakowski, Kiyotaka Fukamachi, Jamshid H Karimov

Cite

APA

Kuroda, T., Miyagi, C., Polakowski, A. R., Flick, C. R., Kuban, B. D., Fukamachi, K., & Karimov, J. H. (2024). Preservation of pulsatility with universal ventricular assist device: In vitro assessment for biventricular support. Artificial Organs. https://doi.org/10.1111/aor.14656

Vancouver

Kuroda T, Miyagi C, Polakowski AR, Flick CR, Kuban BD, Fukamachi K, et al. Preservation of pulsatility with universal ventricular assist device: In vitro assessment for biventricular support. Artificial Organs. 2024. doi:10.1111/aor.14656

BibTeX
@article{Kuroda2024,
  author = {Kuroda, Taiyo and Miyagi, Chihiro and Polakowski, Anthony R and Flick, Christine R and Kuban, Barry D and Fukamachi, Kiyotaka and Karimov, Jamshid H},
  title = {Preservation of pulsatility with universal ventricular assist device: In vitro assessment for biventricular support},
  journal = {Artificial Organs},
  year = {2024},
  doi = {10.1111/aor.14656}
}
Start for free →