Evaluation of ultra‐low‐dose CT with tin filter for craniosynostosis
Journal of Medical Imaging and Radiation Oncology · 2025
DOI: 10.1111/1754-9485.13812 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Medical Imaging and Radiation Oncology, 2025. Verfasst von Wilson Tao und 1 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Children's Hospital Westmead, Sydney, New South Wales, Australia. Hauptthemen laut MEDLINE: Craniosynostoses, Tomography, X-Ray Computed, Radiation Protection. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 23 Literaturangaben.
Abstract (Originalsprache)
Introduction
CT has replaced skull radiography as the gold standard for assessment of craniosynostosis in children. To minimise the risks of ionising radiation in this radiosensitive population, low‐dose CT protocols are increasingly being adopted. This study evaluates the effectiveness of an ultra‐low‐dose CT protocol with a tin filter in reducing radiation exposure whilst maintaining diagnostic quality for craniosynostosis, and its utility in the evaluation of other findings not appreciable on skull radiography.
Methods
Twenty‐seven patients who underwent ultra‐low‐dose CT for craniosynostosis were compared with an age‐matched control group who received standard‐dose CT for indications other than craniosynostosis. Differences in radiation dosimetry and quantitative image quality parameters were analysed using independent two‐tailed t ‐tests. Radiologist reports were also examined for the frequency of other incidental radiological findings.
Results
Mean effective dose for the 27 ultra‐low‐dose CT scans was 0.14 mSv, a 92% reduction compared with the control group. Image quality, measured by contrast‐to‐noise ratio, was significantly lower in the ultra‐low‐dose scans compared with the standard‐dose scans; however, all scans were diagnostic for detecting or excluding craniosynostosis. Hydrocephalus was able to be ruled out in all 27 ultra‐low‐dose scans.
Conclusion
Ultra‐low‐dose CT with a tin filter allows for the diagnosis of craniosynostosis with a 92% dose reduction compared with the standard CT protocol. Despite lower image quality with the ultra‐low‐dose protocol when compared to standard CT, all 27 scans were adequate for craniosynostosis diagnosis. Additionally, it allowed for the assessment of other relevant findings not appreciable with skull radiography, highlighting its advantages as the new clinical standard for craniosynostosis assessment.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Radiology, Children's Hospital Westmead, Sydney, New South Wales, Australia.
- Faculty of Medicine and Health, University of Sydney, Sydney, New South Wales, Australia.
ORCID-Profile: Wilson Tao, Robert Goetti
Zitieren
APA
Tao, W., & Goetti, R. (2025). Evaluation of ultra‐low‐dose CT with tin filter for craniosynostosis. Journal of Medical Imaging and Radiation Oncology. https://doi.org/10.1111/1754-9485.13812
Vancouver
Tao W, Goetti R. Evaluation of ultra‐low‐dose CT with tin filter for craniosynostosis. Journal of Medical Imaging and Radiation Oncology. 2025. doi:10.1111/1754-9485.13812
BibTeX
@article{Tao2025,
author = {Tao, Wilson and Goetti, Robert},
title = {Evaluation of ultra‐low‐dose CT with tin filter for craniosynostosis},
journal = {Journal of Medical Imaging and Radiation Oncology},
year = {2025},
doi = {10.1111/1754-9485.13812}
}
EUVIMED