Randoms Estimation for Long Axial Field-of-View PET
IEEE Transactions on Radiation and Plasma Medical Sciences · 2026-03 · 10 (3) · 401-405
DOI: 10.1109/trpms.2025.3591035 ↗Kurz gefasst
Fachartikel in der Zeitschrift IEEE Transactions on Radiation and Plasma Medical Sciences, 2026-03. Verfasst von Margaret E Daube-Witherspoon und 2 weiteren Autor:innen. Beteiligte Einrichtung: Dept. of Radiology, University of Pennsylvania, Philadelphia, PA 19104 USA. Die Arbeit nennt 19 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NCI NIH HHS, National Institutes of Health.
Abstract (Originalsprache)
The high sensitivity of long axial field-of-view (AFOV) PET scanners has enabled studies over a wide range of count rates and count densities. However, these systems have a large axial acceptance angle that necessitates a wide coincidence window to capture the oblique true coincidences. In addition, the measured delays sinogram is sparse and noisy. We studied four methods of randoms estimation on a long AFOV system to assess their impact on accuracy and image noise: measured delays using a delayed coincidence window (RD), 2D Casey averaging of measured delays (RD-smooth), 2D average of measured delays (RD-ave - the current default method on the PennPET Explorer), and estimation of randoms from singles (RS). We looked at cases with varying count densities, randoms fractions, and non-pure positron emitters. A positive bias observed at low randoms counts for the RD and RD-smooth methods was not seen with the RD-ave or RS methods. For all cases, quantitative results with RS agreed to within 2.5% of the RD-ave method, while RD and RD-smooth estimates showed differences of 5-49%, with larger differences in areas of low uptake. The RS method is a practical technique for list-mode data and list-mode reconstruction by reducing the size of stored list events. It also avoids small approximations in the RD-ave method. For long AFOV systems, estimating randoms from singles is a practical and accurate method.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Dept. of Radiology, University of Pennsylvania, Philadelphia, PA 19104 USA.
Förderung
- NCI NIH HHS
- National Institutes of Health
Zitieren
APA
Daube-Witherspoon, M. E., Moore, S. C., & Karp, J. S. (2026). Randoms Estimation for Long Axial Field-of-View PET. IEEE Transactions on Radiation and Plasma Medical Sciences, 10(3), 401-405. https://doi.org/10.1109/trpms.2025.3591035
Vancouver
Daube-Witherspoon ME, Moore SC, Karp JS. Randoms Estimation for Long Axial Field-of-View PET. IEEE Transactions on Radiation and Plasma Medical Sciences. 2026;10(3):401-405. doi:10.1109/trpms.2025.3591035
BibTeX
@article{DaubeWitherspoon2026,
author = {Daube-Witherspoon, Margaret E and Moore, Stephen C and Karp, Joel S},
title = {Randoms Estimation for Long Axial Field-of-View PET},
journal = {IEEE Transactions on Radiation and Plasma Medical Sciences},
year = {2026},
volume = {10},
number = {3},
pages = {401--405},
doi = {10.1109/trpms.2025.3591035}
}
EUVIMED