Comparison of effects of peripheral vasculature on tonometric radial pulse and cuff-based brachial pulse waveform as used in estimation of central aortic pressures

Mark Butlin, Isabella Tan, Ahmad Qasem, Alberto P Avolio

2023 45th Annual International Conference of the IEEE Engineering in Medicine & Biology Society (EMBC) · 2023

DOI: 10.1109/embc40787.2023.10340973 ↗

Kurz gefasst

Fachartikel in der Zeitschrift 2023 45th Annual International Conference of the IEEE Engineering in Medicine & Biology Society (EMBC), 2023. Verfasst von Mark Butlin und 3 weiteren Autor:innen. Hauptthemen laut MEDLINE: Arterial Pressure, Blood Pressure Determination. Die Arbeit nennt 11 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

OBJECTIVE: Aortic pressure estimation requires reliable peripheral pulse waveform acquisition. The peripheral waveform can change with local vascular effects that can be independent of aortic pressure. This study quantifies the effects of peripheral vasculature changes on radial and brachial waveforms. DESIGN AND METHOD: In 20 subjects (37± 15 years, 7 female), brachial volumetric displacement (cuff-based) and radial tonometry waveforms were simultaneously measured whilst a cuff around the hand on the same arm was inflated to induce transmural pressures of -60, -30, -15, 0, 15 and 30 mmHg, altering local peripheral resistance and compliance by graded arterial wall unloading. Aortic blood pressure (BP), augmentation index (AIx) and ejection duration were calculated from the measurements using a generalized transfer function. The parameters under unloaded conditions were compared to baseline measurements. RESULTS: Brachial systolic and diastolic BP did not change throughout the experiment. Altering peripheral resistance and compliance did not significantly change calculated aortic BP values, although changes were nominally greater for radial (maximum +8±1 mmHg) compared to brachial (maximum +2±1 mmHg) waveforms. AIx at 0 mmHg transmural pressure (maximum arterial wall unloading) was higher when derived from radial waveforms (+24±3%, p

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Arterial PressureBlood Pressure DeterminationHumansFemaleBrachial ArteryBlood PressureManometry

Zitieren

APA

Butlin, M., Tan, I., Qasem, A., & Avolio, A. P. (2023). Comparison of effects of peripheral vasculature on tonometric radial pulse and cuff-based brachial pulse waveform as used in estimation of central aortic pressures. 2023 45th Annual International Conference of the IEEE Engineering in Medicine & Biology Society (EMBC). https://doi.org/10.1109/embc40787.2023.10340973

Vancouver

Butlin M, Tan I, Qasem A, Avolio AP. Comparison of effects of peripheral vasculature on tonometric radial pulse and cuff-based brachial pulse waveform as used in estimation of central aortic pressures. 2023 45th Annual International Conference of the IEEE Engineering in Medicine & Biology Society (EMBC). 2023. doi:10.1109/embc40787.2023.10340973

BibTeX
@article{Butlin2023,
  author = {Butlin, Mark and Tan, Isabella and Qasem, Ahmad and Avolio, Alberto P},
  title = {Comparison of effects of peripheral vasculature on tonometric radial pulse and cuff-based brachial pulse waveform as used in estimation of central aortic pressures},
  journal = {2023 45th Annual International Conference of the IEEE Engineering in Medicine & Biology Society (EMBC)},
  year = {2023},
  doi = {10.1109/embc40787.2023.10340973}
}
Kostenlos starten →