Assessing the Accuracy of Systolic Aortic Pressure Estimation From a Brachial Cuff Alone: A Comparison With Radial Tonometry
American Journal of Hypertension · 2024
DOI: 10.1093/ajh/hpae039 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift American Journal of Hypertension, 2024. Verfasst von Denis Chemla und 5 weiteren Autor:innen. Beteiligte Einrichtung: INSERM UMRS 999, Department of Cardiovascular Physiology, Hôpital Marie Lannelongue, 133 Avenue de la Résistance, 92350 Le Plessis-Robinson, France. Hauptthemen laut MEDLINE: Manometry, Blood Pressure Determination, Arterial Pressure, Brachial Artery. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 20 Literaturangaben.
Abstract (Originalsprache)
BACKGROUND A novel method for estimating central systolic aortic pressure (cSAP) has emerged, relying solely on the peripheral mean (MBP) and diastolic (DBP) blood pressures. We aimed to assess the accuracy of this Direct Central Blood Pressure estimation using cuff alone (DCBPcuff = MBP2/DBP) in comparison to the use of a generalized transfer function to derive cSAP from radial tonometry (cSAPtono). METHODS This retrospective analysis involved the International Database of Central Arterial Properties for Risk Stratification (IDCARS) data (Aparicio et al., Am J Hypertens 2022). The dataset encompassed 10,930 subjects from 13 longitudinal cohort studies worldwide (54.8% women; median age 46.0 years; office hypertension: 40.1%; treated: 61.0%), documenting cSAPtono via SphygmoCor calibrated against brachial systolic BP (SBP) and DBP. Our analysis focused on aggregate group data from 12/13 studies (89% patients) where a full BP dataset was available. A 35% form factor was used to estimate MBP = (DBP + (0.35 × (SBP-DBP)), from which DCBPcuff was derived. The predefined acceptable error for cSAPtono estimation was set at ≤ 5 mm Hg. RESULTS The cSAPtono values ranged from 103.8–127.0 mm Hg (n = 12). The error between DCBPcuff and cSAPtono was 0.2 ± 1.4 mm Hg, with no influence of the mean. Errors ranged from −1.8 to 2.9 mm Hg across studies. No significant difference in errors was observed between BP measurements obtained via oscillometry (n = 9) vs. auscultation (n = 3) (P = 0.50). CONCLUSIONS Using published aggregate group data and a 35% form factor, DCBPcuff demonstrated remarkable accuracy in estimating cSAPtono, regardless of the BP measurement technique. However, given that individual BP values were unavailable, further documentation is required to establish DCBPcuff’s precision.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- INSERM UMRS 999, Department of Cardiovascular Physiology, Hôpital Marie Lannelongue, 133 Avenue de la Résistance, 92350 Le Plessis-Robinson, France.
- Cardiovascular Internal Medicine, Department of Clinical and Surgical Sciences, University of Bologna, BolognaItaly.
- Cardiovascular Medicine Unit, Department of Cardiac-Thoracic-Vascular Diseases, IRCCS AOU of Bologna, Bologna, Italy.
- Department of Otolaryngology and Head and Neck Surgery, Shaare-Zedek Medical Center, Faculty of Medicine, Hebrew University of Jerusalem, Jerusalem, Israel.
- Pulse Wave Consulting, Cardiovascular Department, Saint Leu La Foret 95320France.
- Research Center for Epidemiology and Biostatistics, Sorbonne Paris Cité (CRESS), Université Paris Cité, AP-HP, Diagnosis and Therapeutic Center, Hôtel-Dieu, 1, Place du Parvis Notre Dame, Paris 75004, France.
ORCID-Profile: Denis Chemla
Zitieren
APA
Chemla, D., Agnoletti, D., Attal, P., Millasseau, S., Blacher, J., & Jozwiak, M. (2024). Assessing the Accuracy of Systolic Aortic Pressure Estimation From a Brachial Cuff Alone: A Comparison With Radial Tonometry. American Journal of Hypertension. https://doi.org/10.1093/ajh/hpae039
Vancouver
Chemla D, Agnoletti D, Attal P, Millasseau S, Blacher J, Jozwiak M. Assessing the Accuracy of Systolic Aortic Pressure Estimation From a Brachial Cuff Alone: A Comparison With Radial Tonometry. American Journal of Hypertension. 2024. doi:10.1093/ajh/hpae039
BibTeX
@article{Chemla2024,
author = {Chemla, Denis and Agnoletti, Davide and Attal, Pierre and Millasseau, Sandrine and Blacher, Jacques and Jozwiak, Mathieu},
title = {Assessing the Accuracy of Systolic Aortic Pressure Estimation From a Brachial Cuff Alone: A Comparison With Radial Tonometry},
journal = {American Journal of Hypertension},
year = {2024},
doi = {10.1093/ajh/hpae039}
}
EUVIMED