EUVIMEDEUVIMED entdecken
Crossref

A multi-study analysis of variability in calibration coefficients for radiation dose estimation via the dicentric chromosome assay

Morteza Kosarnia iD, Elham Khakshor, Fatemeh Najafian, Hosein Azimian iD

Radiation Protection Dosimetry · 2026

DOI: 10.1093/rpd/ncag065 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Radiation Protection Dosimetry, 2026. Verfasst von Morteza Kosarnia und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Physics, Faculty of Medicine, Mashhad University of Medical Sciences, Razavi Khorasan Province, Pardis Daneshgah, Vakilabad Blv…. Hauptthemen laut MEDLINE: Chromosome Aberrations, Radiometry, Biological Assay. Die Arbeit nennt 23 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Accurate calibration of biological dosimetry methods is crucial for reliable radiation dose assessment in both medical and research contexts. However, discrepancies in calibration coefficients (α, β, c) across various studies impede standardization and the comparability of dose–response curves. This study conducted a systematic analysis of calibration data from 62 studies covering different radiation types (60Co gamma rays, 137Cs gamma rays, and X-rays) with varying energies and dose rates (0.00196–23.85 Gy/min), to explore the sources of variability and their effects on dicentric chromosome yields. Using descriptive statistics, analysis of variance, Kruskal–Wallis tests, and correlation analyses, we found significant differences in α and β coefficients across different dose rate categories for 60Co gamma rays (P < .05). Specifically, the α coefficient increased from 0.022 ± 0.015 at low dose rates (0–0.5 Gy/min) to 0.049 ± 0.021 at higher dose rates (>1 Gy/min). Moreover, no significant differences were detected between the Computer-Assisted Brain Analysis System (CABAS) and Dose Estimate software for estimating coefficients (P > .05 for α, β, c), indicating consistency in the methodological outputs of these software tools. The observed variability was attributed to factors such as radiation quality, experimental protocols, and inter-laboratory differences. These findings highlight the necessity for standardized calibration protocols that take radiation parameters into account. We propose the development of harmonized experimental setups and reference curves to improve comparability across studies, ultimately enhancing the accuracy of biological dosimetry for radiation protection and epidemiological research. In alignment with existing International Atomic Energy Agency and International Organization for Standardization (ISO) guidelines and acknowledging that international ring-trial intercomparisons remain essential for ensuring comparability across laboratories.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Chromosome AberrationsRadiometryBiological AssayCalibrationHumansGamma RaysCobalt RadioisotopesRadiation DosageDose-Response Relationship, RadiationCesium RadioisotopesX-RaysReproducibility of ResultsSoftware

Substanzen

Cobalt RadioisotopesCobalt-60Cesium RadioisotopesCesium-137

Autor:innen und Institutionen

  • Department of Medical Physics, Faculty of Medicine, Mashhad University of Medical Sciences, Razavi Khorasan Province, Pardis Daneshgah, Vakilabad Blvd., Mashhad 9177948564, Iran.
  • Cellular and Molecular Research Center, Sabzevar University of Medical Sciences, Razavi Khorasan Province, Shohadaye Hasteiee Blvd., Sabzevar 9617913112, Iran.
  • Medical Physics Research Center, Basic Sciences Research Institute, Mashhad University of Medical Sciences, Razavi Khorasan Province, Pardis Daneshgah, Vakilabad Blvd., Mashhad 9177948564, Iran.

ORCID-Profile: Morteza Kosarnia, Hosein Azimian

Zitieren

APA

Kosarnia, M., Khakshor, E., Najafian, F., & Azimian, H. (2026). A multi-study analysis of variability in calibration coefficients for radiation dose estimation via the dicentric chromosome assay. Radiation Protection Dosimetry. https://doi.org/10.1093/rpd/ncag065

Vancouver

Kosarnia M, Khakshor E, Najafian F, Azimian H. A multi-study analysis of variability in calibration coefficients for radiation dose estimation via the dicentric chromosome assay. Radiation Protection Dosimetry. 2026. doi:10.1093/rpd/ncag065

BibTeX
@article{Kosarnia2026,
  author = {Kosarnia, Morteza and Khakshor, Elham and Najafian, Fatemeh and Azimian, Hosein},
  title = {A multi-study analysis of variability in calibration coefficients for radiation dose estimation via the dicentric chromosome assay},
  journal = {Radiation Protection Dosimetry},
  year = {2026},
  doi = {10.1093/rpd/ncag065}
}
Kostenlos starten →