Results of the proficiency test for the Ka CMC of the SSDL-ININ Mexico, for diagnostic radiology IEC 61267:2005 and radiation protection ISO 4037:2019 X-ray beam qualities
Radiation Protection Dosimetry · 2024
DOI: 10.1093/rpd/ncae208 ↗Kurz gefasst
Fachartikel in der Zeitschrift Radiation Protection Dosimetry, 2024. Verfasst von José Trinidad Álvarez Romero und 4 weiteren Autor:innen. Beteiligte Einrichtung: Secondary Standard Dosimetry Laboratory, Ionizing Radiation Metrology Department, Instituto Nacional de Investigaciones Nucleares, Carretera Federal …. Hauptthemen laut MEDLINE: Radiation Protection. Die Arbeit nennt 29 Literaturangaben.
Abstract (Originalsprache)
The present work describes the results for the bilateral comparison between the Secondary Standards Dosimetry Laboratories of the Instituto Nacional de Investigaciones Nucleares (SSDL-ININ), Mexico, and the pilot laboratory, the Centro de Protección e Higiene de las Radiaciones (SSDL-CPHR), Cuba, for the realization of the air kerma (${\boldsymbol{K}}_{\boldsymbol{a}}$) quantity to: (i) diagnostic X-ray (DXR) IEC 61267:2005 beam qualities: RQR 5 RQR 9, RQT 8, and RQT 9 and (ii) radiation protection (RP) ISO 4037:2019 X-ray beam qualities: N60, N100, and W60, as well S-Cs for gamma radiation. The matched and characterized reference fields are realized at SSDL-ININ with a Toshiba E7252FX X-ray tube. In all cases, the ${\boldsymbol{K}}_{\boldsymbol{a}}$ measurements are traceable to PTB, except for S-Cs, traceable to ININ. The SSDL-CPHR realizes its reference radiation beam qualities also as matched reference fields with a Pantak HF160C X-ray tube. For the S-Cs beam quality, the CPHR uses a Buchler OB6 irradiator; the ININ, a Siemens Cesagammatron. The transfer ionization chambers calibrated in the RQR and RQT qualities are an Exradin A650/DO70312; for the RQT qualities for CT, a PTW 30009/0655 and RC3CT/9217 were calibrated; and for the RP X-ray qualities, the Exradin A650/DO70312 and a PTW 32002/073 were calibrated. The ratio ${\boldsymbol{R}}_{\boldsymbol{k}}$ of the resulting calibration coefficients ${\boldsymbol{N}}_{{\boldsymbol{K}}_{\boldsymbol{a}}}$ of each SSDL of the RQR qualities are consistent within the $\boldsymbol{U}\left(\boldsymbol{k}=\mathbf{2}\right)\%$ ≤ 2.3%; for the RQT qualities, the differences are less than the $\boldsymbol{U}\left(\boldsymbol{k}=\mathbf{2}\right)\%$ ≤ 3.6%; and for the RP X-ray qualities, within the order of $\boldsymbol{U}\left(\boldsymbol{k}=\mathbf{2}\right)\%$ ≤ 3.5%. For the S-Cs quality, the ${\boldsymbol{R}}_{\boldsymbol{k}}$ is in the order of $\boldsymbol{U}\left(\boldsymbol{k}=\mathbf{2}\right)$ ≤ 3.0%. It should be noted that the ININ U's are generally larger due to the failure of the ${\boldsymbol{k}}_{\boldsymbol{TP}}$ correction factor, a brief discussion is given here. The ${\boldsymbol{E}}_{\boldsymbol{n}}$ score from ISO 17043 provides statistical support to this comparison.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Secondary Standard Dosimetry Laboratory, Ionizing Radiation Metrology Department, Instituto Nacional de Investigaciones Nucleares, Carretera Federal México Toluca S/N, La Marquesa, Ocoyoacac, Estado de México 52750, Mex…
- Secondary Standard Dosimetry Laboratory, Centro de Proteccion e Higiene de las Radiaciones, Carretera La Victoria II Km 2½, e/Monumental y Final, Guanabacoa, La Habana 11100, Cuba.
- Former staff of Secondary Standard Dosimetry Laboratory, Centro de Protección e Higiene de la Radiaciones, Carretera La Victoria II Km 2½, e/Monumental y Final, Guanabacoa, La Habana, 11100, Cuba.
ORCID-Profile: José Trinidad Álvarez Romero
Zitieren
APA
Romero, J. T. Á., Salas, G. W., de Los Angeles Montes Rodríguez, M., Vertti, M. R. C., & Acosta, A. L. R. (2024). Results of the proficiency test for the Ka CMC of the SSDL-ININ Mexico, for diagnostic radiology IEC 61267:2005 and radiation protection ISO 4037:2019 X-ray beam qualities. Radiation Protection Dosimetry. https://doi.org/10.1093/rpd/ncae208
Vancouver
Romero JTÁ, Salas GW, de Los Angeles Montes Rodríguez M, Vertti MRC, Acosta ALR. Results of the proficiency test for the Ka CMC of the SSDL-ININ Mexico, for diagnostic radiology IEC 61267:2005 and radiation protection ISO 4037:2019 X-ray beam qualities. Radiation Protection Dosimetry. 2024. doi:10.1093/rpd/ncae208
BibTeX
@article{Romero2024,
author = {Romero, José Trinidad Álvarez and Salas, Gonzalo Walwyn and de Los Angeles Montes Rodríguez, María and Vertti, Mario Raymundo Cabrera and Acosta, Andy Luis Romero},
title = {Results of the proficiency test for the Ka CMC of the SSDL-ININ Mexico, for diagnostic radiology IEC 61267:2005 and radiation protection ISO 4037:2019 X-ray beam qualities},
journal = {Radiation Protection Dosimetry},
year = {2024},
doi = {10.1093/rpd/ncae208}
}
EUVIMED