Assessment of occupational radiation exposure during a specific endoscopic retrograde cholangiopancreatography procedure

Ahmed Jibril Abdi iD, Malene Roland Vils Pedersen, Helle Precht iD, Karen Brage, Pernille Lund Hansen

Radiation Protection Dosimetry · 2024

DOI: 10.1093/rpd/ncae197 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Radiation Protection Dosimetry, 2024. Verfasst von Ahmed Jibril Abdi und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Clinical Engineering, Region of Southern Denmark, Odense University Hospital, J.B. Winsløws Vej 4, 5000 Odense C, Denmark. Hauptthemen laut MEDLINE: Occupational Exposure, Cholangiopancreatography, Endoscopic Retrograde, Radiation Exposure, Radiation Dosage. Die Arbeit nennt 31 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Aim

The aim of this study was to evaluate the occupational radiation exposure of staff during endoscopic retrograde cholangiopancreatography (ERCP), with a focus on individuals closest to the radiation source, and to identify potential increases in exposure to hands and eye lenses. Patient radiation exposure during ERCP was also assessed. Materials and methods Staff organ doses were monitored using a Philips Allura Xper FD 20 fluoroscopy system, during 24 ERCP procedures for a period of 7 months. Staff doses were measured using thermoluminescence dosemeters and electronic personal dosemeters, and patient effective doses were simulated and calculated.

Results

Physicians’ annual organ doses ranged from 0.2 to 1.6 mSv for shoulders, 0.1 to 0.4 mSv for eye lenses, and 0.3 to 1.6 mSv for fingers. The annual organ dose of the nursing staff ranged from 0.08 to 2.4 mSv for shoulders, 0.02 to 2.3 mSv for eye lenses, and 1.2 to 5.3 mSv for fingers. The effective dose to patients ranged from 0.009 to 0.46 mSv.

Conclusion

Staff doses were within safe limits, but patient doses were high, emphasizing the need for improved radiation protection.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Occupational ExposureCholangiopancreatography, Endoscopic RetrogradeRadiation ExposureRadiation DosageRadiation ProtectionRadiation MonitoringHumansFluoroscopyThermoluminescent DosimetryMaleLens, CrystallineFemale

Autor:innen und Institutionen

  • Department of Clinical Engineering, Region of Southern Denmark, Odense University Hospital, J.B. Winsløws Vej 4, 5000 Odense C, Denmark.
  • Research and Innovation Unit of Radiology, University of Southern Denmark, Kløvervænget 10, 5000 Odense C, Denmark.
  • Department of Regional Health Research, University of Southern Denmark, Campusvej 55, 5230 Odense M, Denmark.
  • Department of Radiology, Vejle Hospital - Part of Lillebaelt Hospital, Beriderbakken 4, 7100 Vejle, Denmark.
  • Health Sciences Research Centre, UCL University College, Niels Bohrs Allé 1, 5230 Odense M, Denmark.
  • Department of Radiology, Lillebaelt Hospital, University Hospitals of Southern Denmark, Sygehusvej 24, 6000 Kolding, Denmark.

ORCID-Profile: Ahmed Jibril Abdi, Helle Precht

Zitieren

APA

Abdi, A. J., Pedersen, M. R. V., Precht, H., Brage, K., & Lund Hansen, P. (2024). Assessment of occupational radiation exposure during a specific endoscopic retrograde cholangiopancreatography procedure. Radiation Protection Dosimetry. https://doi.org/10.1093/rpd/ncae197

Vancouver

Abdi AJ, Pedersen MRV, Precht H, Brage K, Lund Hansen P. Assessment of occupational radiation exposure during a specific endoscopic retrograde cholangiopancreatography procedure. Radiation Protection Dosimetry. 2024. doi:10.1093/rpd/ncae197

BibTeX
@article{Abdi2024,
  author = {Abdi, Ahmed Jibril and Pedersen, Malene Roland Vils and Precht, Helle and Brage, Karen and Lund Hansen, Pernille},
  title = {Assessment of occupational radiation exposure during a specific endoscopic retrograde cholangiopancreatography procedure},
  journal = {Radiation Protection Dosimetry},
  year = {2024},
  doi = {10.1093/rpd/ncae197}
}
Kostenlos starten →