Anna-Leena Manninen iD, Jukka Järvinen iD, Joanna Sierpowska, Timo J Mäkelä, Tuomas T Rissanen

Radiation Protection Dosimetry · 2024

DOI: 10.1093/rpd/ncae182 ↗

Kurz gefasst

Fallbericht in der Zeitschrift Radiation Protection Dosimetry, 2024. Verfasst von Anna-Leena Manninen und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Diagnostic Radiology, Oulu University Hospital, PO Box 10, FI-90029 Oulu, Finland. Hauptthemen laut MEDLINE: Fetus, Monte Carlo Method, Radiation Dosage, Phantoms, Imaging. Die Arbeit nennt 24 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The benefits of a cardiac resynchronization therapy (CRT) implantation are known in severe heart failure and its implantation may also be considered during pregnancy to ensure safe pregnancy and delivery for both the mother and the fetus. This study on a patient case aimed to estimate the absorbed dose (AD) to the fetus during the CRT implantation of 21st week of the pregnant woman. AD measurement was done using anthropomorphic phantom, radiophotoluminescence dosemeters and Monte Carlo simulation utilizing kerma area product (KAP) of the procedures. The measured AD for a phantom’s uterus and heart was 0.116 and 14.7 mGy with total KAP of 27.9 Gycm2. The estimated actual AD to the fetus and the mother’s heart was 0.004 and 1.2 mGy, with total KAP of 1.5 Gycm2. The dose to the fetus can be minimized with the optimized procedure.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

FetusMonte Carlo MethodRadiation DosagePhantoms, ImagingHumansFemalePregnancyAdultPacemaker, ArtificialCardiac Resynchronization TherapyHeart Failure

Autor:innen und Institutionen

  • Department of Diagnostic Radiology, Oulu University Hospital, PO Box 10, FI-90029 Oulu, Finland.
  • Turku Heart Center, Department of Cardiology, Turku University Hospital, PO Box 52, 20521 Turku, Finland.
  • Medical Imaging Centre of Southwest Finland, Department of Radiology, Turku University Hospital, PO Box 52, 20521 Turku, Finland.
  • Department of Medical Physics, Turku University Hospital, PO Box 52, 20521 Turku, Finland.
  • Department of Radiology, Central Hospital of North Karelia, FI-80210 Joensuu, Finland.
  • OYS Heart Center, Oulu University Hospital, Pohde, PO Box 10, FI-90029 Oulu, Finland.

ORCID-Profile: Anna-Leena Manninen, Jukka Järvinen

Zitieren

APA

Manninen, A. L., Järvinen, J., Sierpowska, J., Mäkelä, T. J., & Rissanen, T. T. (2024). Implantation of CRT pacemaker during pregnancy: estimated fetal radiation dose. Radiation Protection Dosimetry. https://doi.org/10.1093/rpd/ncae182

Vancouver

Manninen AL, Järvinen J, Sierpowska J, Mäkelä TJ, Rissanen TT. Implantation of CRT pacemaker during pregnancy: estimated fetal radiation dose. Radiation Protection Dosimetry. 2024. doi:10.1093/rpd/ncae182

BibTeX
@article{Manninen2024,
  author = {Manninen, Anna-Leena and Järvinen, Jukka and Sierpowska, Joanna and Mäkelä, Timo J and Rissanen, Tuomas T},
  title = {Implantation of CRT pacemaker during pregnancy: estimated fetal radiation dose},
  journal = {Radiation Protection Dosimetry},
  year = {2024},
  doi = {10.1093/rpd/ncae182}
}
Kostenlos starten →