EUVIMEDEUVIMED entdecken
ValidierungsstudieOpen Access · CC BY 4.0

Validation of Monte Carlo simulations with measurement data for fetal dose assessment in proton therapy for breast cancer

Jana Hohmann iD, Marijke De Saint-Hubert iD, Menke Weessies, Anneleen Goedgebeur iD, Aude Geeraerts iD, Dries Colson iD, Murillo Bellezzo iD, Gloria Vilches-Freixas iD, Frédéric Amant iD, Maarten Lambrecht iD, Tom Depuydt

Physics in Medicine & Biology · 2026

DOI: 10.1088/1361-6560/ae2aa0 ↗

Kurz gefasst

Validierungsstudie in der Zeitschrift Physics in Medicine & Biology, 2026. Verfasst von Jana Hohmann und 10 weiteren Autor:innen. Beteiligte Einrichtung: Department of Oncology-Laboratory of Experimental Radiotherapy, KU Leuven, Leuven, Belgium. Hauptthemen laut MEDLINE: Monte Carlo Method, Proton Therapy, Breast Neoplasms, Fetus. Die Arbeit nennt 37 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Stichting Tegen Kanker.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective . This study aims to validate a Monte Carlo model for fetal dose estimation in the complex secondary field of pencil beam scanning (PBS) proton therapy for breast cancer, one of the most common cancers occurring during pregnancy. Approach . A TOPAS/GEANT4 Monte Carlo simulation environment based on an IBA ProteusOne beam model was developed, reflecting the experimental setup of a breast irradiation using a pregnant anthropomorphic phantom. Experimental doses were acquired with thermoluminescent dosimeters for protons and gammas, and bubble detectors (BDs) for neutrons. Simulated doses were scored at the same positions using three hadronic physics models: BIC_HP, BIC_AllHP, and BERT_HP. Experimental doses were corrected for detector energy response using simulation-derived energy spectra. Main results . Agreement between simulation and measurement varied depending on hadronic model, scoring volume size, and correcting for BD energy response. Two physics models conservatively estimated fetal neutron doses within the combined measurement and simulation uncertainties, with BIC_AllHP showing the closest agreement. Combined proton and gamma doses were accurately reproduced for all models for inserts 2–6, but were underestimated for insert 1, likely due to dose gradients and modeling limitations near the treatment field. The total simulated fetal dose equivalent at the fundus height was 5.17 mSv. This value is substantially lower than doses reported for photon-based therapies, remains well below the 100 mSv threshold for deterministic effects, and is within range of the public 1 mSv dose limit. Significance . The results demonstrate that, within the tested experimental framework, the TOPAS/GEANT4 Monte Carlo model is suitable for fetal dose estimation in PBS proton therapy for breast cancer. In this setting, calculated fetal doses were substantially lower than those reported for photon-based radiotherapy. The validated framework provides a practical basis for treatment planning optimization and risk assessment and can be extended to other clinical scenarios following similar validation.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Monte Carlo MethodProton TherapyBreast NeoplasmsFetusRadiation DosageHumansFemaleRadiotherapy DosagePregnancyPhantoms, ImagingRadiometry

Stichwörter der Autor:innen

Monte CarloTOPASbreast cancerfetal doseout-of-field-dosepregnancyproton therapy

Autor:innen und Institutionen

  • Department of Oncology-Laboratory of Experimental Radiotherapy, KU Leuven, Leuven, Belgium.
  • Radiation Protection Dosimetry and Calibration Expert Group, Belgian Nuclear Research Center (SCK CEN), Mol, Belgium.
  • Department of Radiation Oncology (Maastro), GROW School for Oncology and Reproduction, Maastricht University Medical Center, Maastricht, The Netherlands.
  • Department of Radiation Oncology, UZ Leuven, Leuven, Belgium.
  • Department of Oncology, Unit of Gynaecological Oncology, KU Leuven, Leuven, Belgium.
  • Center for Gynecological Oncology, Netherlands Cancer Institute, Amsterdam, The Netherlands.

ORCID-Profile: Jana Hohmann, Marijke De Saint-Hubert, Anneleen Goedgebeur, Aude Geeraerts, Dries Colson, Murillo Bellezzo, Gloria Vilches-Freixas, Frédéric Amant, Maarten Lambrecht

Förderung

  • Stichting Tegen Kanker

Zitieren

APA

Hohmann, J., De Saint-Hubert, M., Weessies, M., Goedgebeur, A., Geeraerts, A., Colson, D., Bellezzo, M., Vilches-Freixas, G., Amant, F., Lambrecht, M., & Depuydt, T. (2026). Validation of Monte Carlo simulations with measurement data for fetal dose assessment in proton therapy for breast cancer. Physics in Medicine & Biology. https://doi.org/10.1088/1361-6560/ae2aa0

Vancouver

Hohmann J, De Saint-Hubert M, Weessies M, Goedgebeur A, Geeraerts A, Colson D, et al. Validation of Monte Carlo simulations with measurement data for fetal dose assessment in proton therapy for breast cancer. Physics in Medicine & Biology. 2026. doi:10.1088/1361-6560/ae2aa0

BibTeX
@article{Hohmann2026,
  author = {Hohmann, Jana and De Saint-Hubert, Marijke and Weessies, Menke and Goedgebeur, Anneleen and Geeraerts, Aude and Colson, Dries and Bellezzo, Murillo and Vilches-Freixas, Gloria and Amant, Frédéric and Lambrecht, Maarten and Depuydt, Tom},
  title = {Validation of Monte Carlo simulations with measurement data for fetal dose assessment in proton therapy for breast cancer},
  journal = {Physics in Medicine & Biology},
  year = {2026},
  doi = {10.1088/1361-6560/ae2aa0}
}
Kostenlos starten →