EUVIMEDEUVIMED entdecken
Crossref

Development of α-ray visualization survey meter in high gamma and neutron background environment

Youichi Tsubota iD, Kenji Kobayashi, Tatsuya Ishii, Misaki Hirato, Satoshi Shioya, Takahiro Nakagawa

Radiation Protection Dosimetry · 2024

DOI: 10.1093/rpd/ncae169 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Radiation Protection Dosimetry, 2024. Verfasst von Youichi Tsubota und 5 weiteren Autor:innen. Beteiligte Einrichtung: Collaborative Laboratories for Advanced Decommissioning Science, Japan Atomic Energy Agency, 2-4 Shirakata, Tokai-mura, Ibaraki 319-1195, Japan. Hauptthemen laut MEDLINE: Gamma Rays, Neutrons, Scintillation Counting, Radiation Monitoring. Die Arbeit nennt 8 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

A survey meter was developed to reliably detect and visualize surface contamination of suits and objects by α-nuclides in high γ/n-rays background radiation environment. The survey meter features a semi-opaque ZnS:Ag scintillator mounted directly onto a multi-anode photomultiplier tube (MA-PMT) and amplification circuits, ensuring output gain equalization for all channels. α-ray events induce localized light emission in thin-film scintillators. By directly mounting the scintillator, diffusion of light before reaching the MA-PMT is suppressed, concentrating it in just a few channels, thereby facilitating discrimination from background radiation. This design also enables clear visualization of the shape of surface contamination. The prototyped survey meter is capable of responding up to 2.1 × 107 cpm, with no γ-ray response even in high-radiation environments exceeding 1 Sv/h. In actual environments with high background radiation, contamination of ~1/100th of the surface contamination density limit of 4 Bq/cm2 could be reliably detected.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Gamma RaysNeutronsScintillation CountingRadiation MonitoringBackground RadiationAlpha ParticlesEquipment Design

Autor:innen und Institutionen

  • Collaborative Laboratories for Advanced Decommissioning Science, Japan Atomic Energy Agency, 2-4 Shirakata, Tokai-mura, Ibaraki 319-1195, Japan.
  • Radiation Protection Department, Nuclear Fuel Cycle Engineering Laboratories, Japan Atomic Energy Agency, 4-33 Muramatsu, Tokai-mura, Ibaraki 319-1194, Japan.

ORCID-Profile: Youichi Tsubota

Zitieren

APA

Tsubota, Y., Kobayashi, K., Ishii, T., Hirato, M., Shioya, S., & Nakagawa, T. (2024). Development of α-ray visualization survey meter in high gamma and neutron background environment. Radiation Protection Dosimetry. https://doi.org/10.1093/rpd/ncae169

Vancouver

Tsubota Y, Kobayashi K, Ishii T, Hirato M, Shioya S, Nakagawa T. Development of α-ray visualization survey meter in high gamma and neutron background environment. Radiation Protection Dosimetry. 2024. doi:10.1093/rpd/ncae169

BibTeX
@article{Tsubota2024,
  author = {Tsubota, Youichi and Kobayashi, Kenji and Ishii, Tatsuya and Hirato, Misaki and Shioya, Satoshi and Nakagawa, Takahiro},
  title = {Development of α-ray visualization survey meter in high gamma and neutron background environment},
  journal = {Radiation Protection Dosimetry},
  year = {2024},
  doi = {10.1093/rpd/ncae169}
}
Kostenlos starten →