Estimation of 241Am and 239Pu activity embedded in the tissue using portable planar HPGe detector

Manohari Murugan iD, Sugumar Ponraj, Mathiyarasu Ramasamy, Ponraju Durairaj, Venkatraman Balasubramanian

Radiation Protection Dosimetry · 2024

DOI: 10.1093/rpd/ncae105 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Radiation Protection Dosimetry, 2024. Verfasst von Manohari Murugan und 4 weiteren Autor:innen. Beteiligte Einrichtung: Safety, Quality and Resource Management Group, Indira Gandhi Centre for Atomic Research, Kalpakkam 603012, Tamilnadu, India. Hauptthemen laut MEDLINE: Plutonium, Americium, Germanium. Die Arbeit nennt 25 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This paper describes a procedure for the estimation of 241Am and 239Pu activity present in the human tissue by measuring the depth of contaminant using a portable Planar High Purity Germanium detector (HPGe). The ratios of photopeak counts of X-rays or gammas obtained with the detector coupled to collimator are calculated for the estimation of depth of the contaminant and the optimum one is determined. Since Minimum Detectable Activities (MDA) for the detector coupled to a collimator are higher than that of bare detector, activity must be estimated using bare detector, after locating the contaminant. Two methods are described for the estimation of plutonium coexisting with 241Am: (i) Abundance and isotopic correction for 239Pu (ii) and 239Pu:241Am ratio. The procedure to estimate 239Pu when plutonium isotopes alone are present is also established. An optimum monitoring period to detect the minimum value of intake for both radionuclides corresponding to chelation therapy and excision is also derived.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

PlutoniumAmericiumGermaniumHumansRadiation MonitoringRadiation DosageRadiometry

Substanzen

PlutoniumAmericiumAmericium-241GermaniumPlutonium-239

Autor:innen und Institutionen

  • Safety, Quality and Resource Management Group, Indira Gandhi Centre for Atomic Research, Kalpakkam 603012, Tamilnadu, India.

ORCID-Profile: Manohari Murugan

Zitieren

APA

Murugan, M., Ponraj, S., Ramasamy, M., Durairaj, P., & Balasubramanian, V. (2024). Estimation of 241Am and 239Pu activity embedded in the tissue using portable planar HPGe detector. Radiation Protection Dosimetry. https://doi.org/10.1093/rpd/ncae105

Vancouver

Murugan M, Ponraj S, Ramasamy M, Durairaj P, Balasubramanian V. Estimation of 241Am and 239Pu activity embedded in the tissue using portable planar HPGe detector. Radiation Protection Dosimetry. 2024. doi:10.1093/rpd/ncae105

BibTeX
@article{Murugan2024,
  author = {Murugan, Manohari and Ponraj, Sugumar and Ramasamy, Mathiyarasu and Durairaj, Ponraju and Balasubramanian, Venkatraman},
  title = {Estimation of 241Am and 239Pu activity embedded in the tissue using portable planar HPGe detector},
  journal = {Radiation Protection Dosimetry},
  year = {2024},
  doi = {10.1093/rpd/ncae105}
}
Kostenlos starten →