EUVIMEDEUVIMED entdecken
Crossref

The effect of contact radiation shielding on breast dose during CT abdomen–pelvis: a phantom study

Maria Long, Salman Albeshan iD, Yazeed Alashban iD, Andrew England, Niamh Moore, Rena Young, Paul Bezzina, Mark F McEntee

Radiation Protection Dosimetry · 2023

DOI: 10.1093/rpd/ncad218 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Radiation Protection Dosimetry, 2023. Verfasst von Maria Long und 7 weiteren Autor:innen. Beteiligte Einrichtung: Medical Imaging and Radiation Therapy Department, School of Medicine, UG Assert, Brookfield Health Sciences, University College Cork, Cork T12 AK54, …. Hauptthemen laut MEDLINE: Radiation Protection, Tomography, X-Ray Computed, Phantoms, Imaging, Breast. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 49 Literaturangaben. Gefördert von: King Saud University.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This study aims to investigate if contact shielding reduces breast radiation dose during computed tomography (CT) abdomen–pelvis examinations using automatic tube current modulation to protect one of the four most radiosensitive organs during CT examinations. Dose measurements were taken with and without contact shielding across the anterior and lateral aspects of the breasts and with and without organ dose modulation (ODM) to quantify achievable dose reductions. Although there are no statistically significant findings, when comparing with and without shielding, the mean breast surface dose was reduced by 0.01 μSv without ODM (1.92–1.91 μSv, p = 0.49) and increased by 0.03 μSv with ODM (1.53–1.56 μSv, p = 0.44). Comparing with and without ODM, the mean breast surface dose was reduced by 0.35 μSv with shielding (1.91–1.56 μSv, p = 0.24) and by 0.39 μSv without shielding (1.92–1.53 μSv, p = 0.17). The addition of contact shielding does not provide significant breast surface radiation dose reduction during CT abdomen–pelvis.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiation ProtectionTomography, X-Ray ComputedPhantoms, ImagingBreastPelvisRadiography, AbdominalAbdomenHumansRadiation DosageFemale

Autor:innen und Institutionen

  • Medical Imaging and Radiation Therapy Department, School of Medicine, UG Assert, Brookfield Health Sciences, University College Cork, Cork T12 AK54, Ireland.
  • Radiological Sciences Department, College of Applied Medical Sciences, King Saud University, PO Box 145111, Riyadh 4545, Saudi Arabia.
  • Department of Radiography, Faculty of Health Sciences, University of Malta, Msida MSD 2080, Malta.

ORCID-Profile: Salman Albeshan, Yazeed Alashban

Förderung

  • King Saud University

Zitieren

APA

Long, M., Albeshan, S., Alashban, Y., England, A., Moore, N., Young, R., Bezzina, P., & McEntee, M. F. (2023). The effect of contact radiation shielding on breast dose during CT abdomen–pelvis: a phantom study. Radiation Protection Dosimetry. https://doi.org/10.1093/rpd/ncad218

Vancouver

Long M, Albeshan S, Alashban Y, England A, Moore N, Young R, et al. The effect of contact radiation shielding on breast dose during CT abdomen–pelvis: a phantom study. Radiation Protection Dosimetry. 2023. doi:10.1093/rpd/ncad218

BibTeX
@article{Long2023,
  author = {Long, Maria and Albeshan, Salman and Alashban, Yazeed and England, Andrew and Moore, Niamh and Young, Rena and Bezzina, Paul and McEntee, Mark F},
  title = {The effect of contact radiation shielding on breast dose during CT abdomen–pelvis: a phantom study},
  journal = {Radiation Protection Dosimetry},
  year = {2023},
  doi = {10.1093/rpd/ncad218}
}
Kostenlos starten →