EURADOS IC2019exteye- a self-funded intercomparison for extremity and eye lens dosemeters organised by EURADOS working group 2
Radiation Protection Dosimetry · 2023
DOI: 10.1093/rpd/ncac275 ↗Kurz gefasst
Fachartikel in der Zeitschrift Radiation Protection Dosimetry, 2023. Verfasst von Andrew McWhan und 5 weiteren Autor:innen. Beteiligte Einrichtung: Charthouse Data Management Limited, Unit A17 Admiralty Park, Poole BH16 6HX, United Kingdom. Hauptthemen laut MEDLINE: Radiation Monitoring, Radiation Protection, Lens, Crystalline, Occupational Exposure. Die Arbeit nennt 11 Literaturangaben.
Abstract (Originalsprache)
The European Radiation Dosimetry Group (EURADOS) has been organising regular, self-funding, external dosimetry intercomparisons (ICs) since 2008. IC2019exteye, for extremity and eye lens dosemeters, attracted 113 systems (68 extremity and 45 eye lens) from 60 institutes with participants from 26 countries around the world. Irradiations were carried out in accredited European irradiation facilities in terms of Hp(0.07)/or Hp(3) in the ranges: photon energy: 16–662 keV, beta mean energy 250–1000 keV, dose: 0.5 mSv to 1 Sv, maximum angle of incidence: ±60°. This paper describes the IC2019exteye set-up, irradiation plan and coordination with an overview of the main conclusions. Out of the total of 113 systems, 57 reported results for photons only and 56 reported both photon and beta. In general, the participants showed very satisfactory performances with the medians of all Hp(0.07) and Hp(3) response values very close to unity.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Charthouse Data Management Limited, Unit A17 Admiralty Park, Poole BH16 6HX, United Kingdom.
- Cavendish Nuclear Limited, A11 Berkeley ADS, Berkeley Centre, Berkeley, Gloucestershire GL13 9FB, United Kingdom.
- CIEMAT, Radiation Dosimetry, Avda Complutense 40, 28040 Madrid, Spain.
- Mirion Dosimetry Services, PO Box 60067, 6800 JB Arnhem, The Netherlands.
- Seibersdorf Labor GmbH, A-2444 Seibersdorf, Austria.
- Mirion Technologies (AWST) GmbH, Auswertungsstelle, Otto-Hahn-Ring 6, D-81739 München, Germany.
ORCID-Profile: Wioletta Dobrzynska, Ana Romero, Hannes Stadtmann
Zitieren
APA
McWhan, A., Dobrzynska, W., Romero, A., Grimbergen, T., Stadtmann, H., & Figel, M. (2023). EURADOS IC2019exteye- a self-funded intercomparison for extremity and eye lens dosemeters organised by EURADOS working group 2. Radiation Protection Dosimetry. https://doi.org/10.1093/rpd/ncac275
Vancouver
McWhan A, Dobrzynska W, Romero A, Grimbergen T, Stadtmann H, Figel M. EURADOS IC2019exteye- a self-funded intercomparison for extremity and eye lens dosemeters organised by EURADOS working group 2. Radiation Protection Dosimetry. 2023. doi:10.1093/rpd/ncac275
BibTeX
@article{McWhan2023,
author = {McWhan, Andrew and Dobrzynska, Wioletta and Romero, Ana and Grimbergen, Tom and Stadtmann, Hannes and Figel, Markus},
title = {EURADOS IC2019exteye- a self-funded intercomparison for extremity and eye lens dosemeters organised by EURADOS working group 2},
journal = {Radiation Protection Dosimetry},
year = {2023},
doi = {10.1093/rpd/ncac275}
}
EUVIMED