Abstract (Originalsprache)
The purpose of this paper is to describe a method for optimizing the calibration period of a dosimetry measurement system with a focus on the personal dose equivalents Hp(d) and individual monitoring services. The systematic and analytical method developed provides a justification to the calibration interval based on experimental data and measurement uncertainties. Its main input is the metrological history of the equipment that is then used to predict the drift of the measurement system. The method is used to make the calibration operation more efficient over time in terms of cost and metrology. After a description of the method itself, the example of the OSL dosemeter reading process will be discussed and results will be shown. The method developed may be applied to other measurement processes no matter the technology used.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Landauer, 9, rue Paul Dautier-CS 60731, 78457 Cedex, Vélizy-Villacoublay, France.
ORCID-Profile: Brahim Moreno
Zitieren
APA
Moreno, B. (2023). Optimization of calibration interval based on equipment metrological history. Radiation Protection Dosimetry. https://doi.org/10.1093/rpd/ncac230
Vancouver
Moreno B. Optimization of calibration interval based on equipment metrological history. Radiation Protection Dosimetry. 2023. doi:10.1093/rpd/ncac230
BibTeX
@article{Moreno2023,
author = {Moreno, Brahim},
title = {Optimization of calibration interval based on equipment metrological history},
journal = {Radiation Protection Dosimetry},
year = {2023},
doi = {10.1093/rpd/ncac230}
}
EUVIMED