Calibration of the W-PIE neutron spectrometer at CERF reference facility
Radiation Protection Dosimetry · 2023
DOI: 10.1093/rpd/ncac203 ↗Kurz gefasst
Fachartikel in der Zeitschrift Radiation Protection Dosimetry, 2023. Verfasst von Andrea Cirillo und 1 weiteren Autor:innen. Beteiligte Einrichtung: Politecnico di Milano - Dipartimento di Energia, Via Labmruschini 4, 20156 Milano, Italy. Hauptthemen laut MEDLINE: Radiation Monitoring. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 7 Literaturangaben.
Abstract (Originalsprache)
The W-PIE is a cosmic neutron spectrometer used for environmental measurements, developed by Politecnico di Milano. The instrument is based on the Artkis M800 thermal neutron detector and works as a 4-channel spectrometer for detecting neutrons in the energy range of 0.01 eV–1 GeV. After the spectrometer had been calibrated in some monoenergetic neutron fields at the PIAF facility of PTB, the instrument was tested in the neutron field generated at the CERF facility, whose spectrum fairly resembles the cosmic one. This measurement is a precious tool for predicting the performance of the instrument in a realistic situation.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Politecnico di Milano - Dipartimento di Energia, Via Labmruschini 4, 20156 Milano, Italy.
ORCID-Profile: Andrea Cirillo, Marco Caresana
Zitieren
APA
Cirillo, A., & Caresana, M. (2023). Calibration of the W-PIE neutron spectrometer at CERF reference facility. Radiation Protection Dosimetry. https://doi.org/10.1093/rpd/ncac203
Vancouver
Cirillo A, Caresana M. Calibration of the W-PIE neutron spectrometer at CERF reference facility. Radiation Protection Dosimetry. 2023. doi:10.1093/rpd/ncac203
BibTeX
@article{Cirillo2023,
author = {Cirillo, Andrea and Caresana, Marco},
title = {Calibration of the W-PIE neutron spectrometer at CERF reference facility},
journal = {Radiation Protection Dosimetry},
year = {2023},
doi = {10.1093/rpd/ncac203}
}
EUVIMED