Establishing multianalyte thresholds for automated detection of EDTA contamination
Laboratory Medicine · 10.06.2026 · 57 (4)
DOI: 10.1093/labmed/lmag041 ↗Kurz gefasst
Fachartikel in der Zeitschrift Laboratory Medicine, online erschienen am 17.07.2026. Verfasst von Jillian Kodger und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Laboratory Medicine, Yale School of Medicine, New Haven, CT, United States. Hauptthemen laut MEDLINE: Edetic Acid, Automation, Laboratory, Blood Chemical Analysis. Die Arbeit nennt 6 Literaturangaben.
Abstract (Originalsprache)
Introduction
Bias in analytes due to EDTA contamination is easy to miss and represents a common and clinically relevant preanalytical error. We designed a study to determine the in vitro effects of rising rates of EDTA contamination in blood samples and to evaluate how these findings could inform the development of multianalyte thresholds for detecting such errors automatically.
Methods
We incrementally spiked blood samples with increasing volumes of EDTA‑contaminated specimen and measured the resulting changes in alkaline phosphatase, calcium, magnesium, potassium, and iron levels. Using these observations, we developed a series of multianalyte thresholds and evaluated their performance against 1 year of retrospective clinical data (n = 2407).
Results
Our experiments demonstrated that increasing rates of EDTA contamination produced changes in several analytes. Accordingly, we applied multianalyte thresholds using potassium and calcium to retrospective samples. Across all samples, smaller magnitudes of analyte deviation resulted in higher detection rates of subtle EDTA contamination but poorer specificity. We selected a rule of a potassium level of 6.5 mmol/L or higher and a calcium level below 6.5 mg/dL (1.63 mmol/L), which identified 14 samples with a contamination rate of 42.8%.
Discussion
Multianalyte thresholds show promise as an effective approach for automatically identifying clinically significant EDTA contamination in clinical specimens.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Laboratory Medicine, Yale School of Medicine, New Haven, CT, United States.
ORCID-Profile: Thomas J S Durant, Joe M El-Khoury
Zitieren
APA
Kodger, J., Durant, T. J. S., Miconi, D., & El-Khoury, J. M. (2026). Establishing multianalyte thresholds for automated detection of EDTA contamination. Laboratory Medicine, 57(4). https://doi.org/10.1093/labmed/lmag041
Vancouver
Kodger J, Durant TJS, Miconi D, El-Khoury JM. Establishing multianalyte thresholds for automated detection of EDTA contamination. Laboratory Medicine. 2026;57(4). doi:10.1093/labmed/lmag041
BibTeX
@article{Kodger2026,
author = {Kodger, Jillian and Durant, Thomas J S and Miconi, Dustin and El-Khoury, Joe M},
title = {Establishing multianalyte thresholds for automated detection of EDTA contamination},
journal = {Laboratory Medicine},
year = {2026},
volume = {57},
number = {4},
doi = {10.1093/labmed/lmag041}
}
EUVIMED