Minimizing serum creatinine interference in automated laboratory analyses through the implementation of urea/creatinine ratio alerting mechanisms
Clinica Chimica Acta · 2025
DOI: 10.1016/j.cca.2025.120190 ↗Kurz gefasst
Fachartikel in der Zeitschrift Clinica Chimica Acta, 2025. Verfasst von Hui Ding und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Laboratory Medicine, Tongren Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai 200336, China. Hauptthemen laut MEDLINE: Creatinine, Automation, Laboratory, Urea, Blood Chemical Analysis. Die Arbeit nennt 23 Literaturangaben.
Abstract (Originalsprache)
BACKGROUND: This study was designed to increase the precision and clinical significance of serum creatinine (Cr) measurements by refining the detection methodology through comprehensive laboratory automation and by employing the urea/creatinine ratio(UCR) to detect and mitigate drug interference. METHODS: We measured the serum Cr concentration via the Siemens sarcosine oxidase enzymatic (SOE) assay. A substantial dataset comprising nearly 2 million records from 98,377 individuals was extracted to form the foundational database for serum UCR analysis. Using the Refine R package, we delineated the reference intervals for UCR. Our laboratory's automated system adeptly identified samples exhibiting aberrant UCR values, prompting a re-evaluation of Cr levels via the Siemens Jaffe (improved alkaline picrate) assay. These Cr measurements were subsequently corroborated against Beckman and Ortho SOE assays and benchmarked against liquid chromatography-isotope dilution tandem mass spectrometry (LC-IDMS/MS), the reference standard. RESULTS: The reference interval for the serum UCR ranged from 0.047 to 0.143(90% CI 0.046-0.048, 0.135-0.148).Among the 58 abnormal samples assessed, the mean deviation in Cr detection for the Siemens, Beckman, and Ortho SOE assays relative to the LC-IDMS/MS results were -61.05%,-53.97%, and -17.18%, respectively. The Siemens Jaffe (improved alkaline picrate) assay revealed a mean deviation of -3.10% compared with that of LC-IDMS/MS. These findings indicate that the improved alkaline picrate assay can accurately reflect the true Cr levels in serum samples affected by calcium dobesilate or etamsylate interference. CONCLUSION: The implementation of an automated UCR-based screening protocol has been shown to efficiently pinpoint samples with drug-related interference. This strategic approach not only decreases clinical risk but also guarantees the fidelity of serum Cr measurements without undermining the overall productivity of the laboratory's automated workflow.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Laboratory Medicine, Tongren Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai 200336, China.
- Department of General Surgery Gastrointestinal Group,Tongren Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai 200336, China.
- Central Labortary ,Shanghai Xuhui Central Hosptial/Zhongshan-Xuhui Hospital, Fudan University, Shanghai 200031,China.
- Department of Laboratory Medicine, Tongren Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai 200336, China. Electronic address: jinpiaolin@163.com.
- Central Labortary ,Shanghai Xuhui Central Hosptial/Zhongshan-Xuhui Hospital, Fudan University, Shanghai 200031,China. Electronic address: shuijunli@163.com.
- Department of Laboratory Medicine, Tongren Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai 200336, China; Department of Laboratory Medicine, College of Health Science and Technology, Shanghai Jiao T…
Zitieren
APA
Ding, H., Shao, J., Cao, H., Xiao, C., Tong, M., Zuo, X., Huang, Z., Lin, J., Li, S., & Sheng, H. (2025). Minimizing serum creatinine interference in automated laboratory analyses through the implementation of urea/creatinine ratio alerting mechanisms. Clinica Chimica Acta. https://doi.org/10.1016/j.cca.2025.120190
Vancouver
Ding H, Shao J, Cao H, Xiao C, Tong M, Zuo X, et al. Minimizing serum creatinine interference in automated laboratory analyses through the implementation of urea/creatinine ratio alerting mechanisms. Clinica Chimica Acta. 2025. doi:10.1016/j.cca.2025.120190
BibTeX
@article{Ding2025,
author = {Ding, Hui and Shao, Jun and Cao, Hua and Xiao, Can and Tong, Minghong and Zuo, Xuemei and Huang, Zhonghua and Lin, Jinpiao and Li, Shuijun and Sheng, Huiming},
title = {Minimizing serum creatinine interference in automated laboratory analyses through the implementation of urea/creatinine ratio alerting mechanisms},
journal = {Clinica Chimica Acta},
year = {2025},
doi = {10.1016/j.cca.2025.120190}
}
EUVIMED