A Gradual Positive Bias in Chloride Measurements with a Major Automated Chemistry Platform

Kwaku Twum, Nichole Korpi-Steiner iD, Ruhan Wei, Steven W Cotten iD

The Journal of Applied Laboratory Medicine · 2025

DOI: 10.1093/jalm/jfaf046 ↗

Kurz gefasst

Fachartikel in der Zeitschrift The Journal of Applied Laboratory Medicine, 2025. Verfasst von Kwaku Twum und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Pathology and Laboratory Medicine, University of North Carolina at Chapel Hill, Chapel Hill, NC, United States. Hauptthemen laut MEDLINE: Chlorides, Automation, Laboratory. Die Arbeit nennt 11 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Chloride is an extracellular anion commonly measured using ion-selective electrode methods standardized to traceable reference materials. This study investigates the gradual appearance of a positive bias in chloride measurements using the Siemens Healthineers Atellica CH platform that was undetectable via routine quality control or proficiency testing.

Methods

The bias was characterized using retrospective analysis of laboratory values from presumable healthy outpatients (ICD-10 code Z00.00), reference interval verifications, peer group means from the College of American Pathologists’ proficiency testing C survey, and a method comparison study with an alternate platform.

Results

Retrospective analysis of patient data (2017–2024) revealed a gradual shift in median patient chloride concentrations distinct from stable trends observed for other electrolytes. In 2024, reference interval verification revealed that 64.3% of chloride results were above the reference interval, even though the verification was successful in 2021. Proficiency testing results were acceptable compared to peer method; however, revealed a mean positive bias of 3.19 mmol/L when compared to other indirect ISE methods (2019–2024). A method comparison with an alternate chemistry platform demonstrated a mean bias of +4.3 mmol/L using remnant patient samples.

Conclusions

The use of peer targets for quality control and proficiency testing concealed the detection of a gradual drift in chloride measurements in our laboratory. Analysis of aggregate patient data using ICD-10 codes represents a powerful tool for laboratories when assay performance is questioned. These findings highlight the need for vigilant monitoring of assay drift even for routine analytes to avert potential impact on patient care.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ChloridesAutomation, LaboratoryHumansRetrospective StudiesQuality ControlLaboratory Proficiency TestingReference ValuesBias

Substanzen

Chlorides

Autor:innen und Institutionen

  • Department of Pathology and Laboratory Medicine, University of North Carolina at Chapel Hill, Chapel Hill, NC, United States.
  • Department of Pathology, Duke University Medical Center, Durham, NC, United States.

ORCID-Profile: Nichole Korpi-Steiner, Steven W Cotten

Zitieren

APA

Twum, K., Korpi-Steiner, N., Wei, R., & Cotten, S. W. (2025). A Gradual Positive Bias in Chloride Measurements with a Major Automated Chemistry Platform. The Journal of Applied Laboratory Medicine. https://doi.org/10.1093/jalm/jfaf046

Vancouver

Twum K, Korpi-Steiner N, Wei R, Cotten SW. A Gradual Positive Bias in Chloride Measurements with a Major Automated Chemistry Platform. The Journal of Applied Laboratory Medicine. 2025. doi:10.1093/jalm/jfaf046

BibTeX
@article{Twum2025,
  author = {Twum, Kwaku and Korpi-Steiner, Nichole and Wei, Ruhan and Cotten, Steven W},
  title = {A Gradual Positive Bias in Chloride Measurements with a Major Automated Chemistry Platform},
  journal = {The Journal of Applied Laboratory Medicine},
  year = {2025},
  doi = {10.1093/jalm/jfaf046}
}
Kostenlos starten →