Development and validation of an liquid chromatography–tandem mass spectrometry method for the detection of substituted phenethylamines and phenidates designated as temporary narcotics in urine
Journal of Analytical Toxicology · 26.09.2026 · 50 (9)
DOI: 10.1093/jat/bkag086 ↗Kurz gefasst
Validierungsstudie in der Zeitschrift Journal of Analytical Toxicology, online erschienen am 22.09.2026. Verfasst von Sujin Jeong und 5 weiteren Autor:innen. Beteiligte Einrichtung: Toxicology and Narcotics Division, Seoul Institute, National Forensic Service, Seoul 08036, South Korea. Hauptthemen laut MEDLINE: Phenethylamines, Substance Abuse Detection, Narcotics, Methylphenidate. Die Arbeit nennt 36 Literaturangaben. Gefördert von: National Forensic Service, Ministry of the Interior and Safety.
Abstract (Originalsprache)
Novel psychoactive phenethylamines and phenidates are synthetically designed analogs of conventional substances such as 3,4-methylenedioxymethamphetamine (MDMA), amphetamine, and methylphenidate, developed to circumvent existing drug regulations. In Korea, the abuse of these substances has increased, particularly among habitual users and newly addicted younger individuals. To address this issue, new substances are regulated through the analog legislation and the temporary narcotic designation system. In this study, liquid chromatography–tandem mass spectrometry method using solid-phase extraction was developed and validated for the simultaneous determination of 13 substituted phenethylamines and 5 substituted phenidates in human urine. The target analytes included 5-MAPDB, BDB, 4-FEA, PMEA, 3F-Phenetrazine, 25iP-NBOMe, 2C-N, 2C-TFM, 30C-NBOMe, 3C-P, 4-MMA-NBOMe, DOI, and 2C-H, as well as 4-MeTMP, 4F-EPH, 4F-MPH, IPPH, and HDEP-28. No interfering peaks were observed at the retention times of analytes and internal standards. Linearity was confirmed over analyte-specific concentration within the overall range of 0.1–100 ng/ml, with coefficient of determination (r2) exceeding 0.995. The limits of detection and quantification were 0.001–0.1 and 0.1–0.2 ng/ml, respectively. Accuracy and precision, expressed as bias and coefficient of variation (CV), were within ±10% and <10%, respectively. The CVs for matrix effects and extraction efficiencies were approximately 10%, and the extraction efficiencies exceeded 70%, confirming the reliability of the method for forensic applications. The validated method enables simultaneous determination of substituted phenethylamines and phenidates in urine and facilitates prompt positive identification in forensic casework. In addition, the structural information, multiple reaction monitoring conditions, and analytical method validation results presented in this study can serve as a basis for expanding research on substances with similar core structures, which is expected to strengthen regulatory responses to evolving drug trends.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Substanzen
Autor:innen und Institutionen
- Toxicology and Narcotics Division, Seoul Institute, National Forensic Service, Seoul 08036, South Korea.
- College of Pharmacy, Chungnam National University, Daejeon 34134, South Korea.
- Jeju Branch, National Forensic Service, Jeju-do 63309, South Korea.
ORCID-Profile: Sujin Jeong, Hyung Min Kim
Förderung
- National Forensic Service
- Ministry of the Interior and Safety
Zitieren
APA
Jeong, S., Kim, Y., Kang, S., Lee, C. E., Jang, M., & Kim, H. M. (2026). Development and validation of an liquid chromatography–tandem mass spectrometry method for the detection of substituted phenethylamines and phenidates designated as temporary narcotics in urine. Journal of Analytical Toxicology, 50(9). https://doi.org/10.1093/jat/bkag086
Vancouver
Jeong S, Kim Y, Kang S, Lee CE, Jang M, Kim HM. Development and validation of an liquid chromatography–tandem mass spectrometry method for the detection of substituted phenethylamines and phenidates designated as temporary narcotics in urine. Journal of Analytical Toxicology. 2026;50(9). doi:10.1093/jat/bkag086
BibTeX
@article{Jeong2026,
author = {Jeong, Sujin and Kim, Yoseob and Kang, Seojin and Lee, Chae Eun and Jang, Moonhee and Kim, Hyung Min},
title = {Development and validation of an liquid chromatography–tandem mass spectrometry method for the detection of substituted phenethylamines and phenidates designated as temporary narcotics in urine},
journal = {Journal of Analytical Toxicology},
year = {2026},
volume = {50},
number = {9},
doi = {10.1093/jat/bkag086}
}