EUVIMEDEUVIMED entdecken
ValidierungsstudieVergleichsstudie

Validation of an LC–MS–MS method for analysis of 58 drugs of abuse in oral fluid and method comparison with an established LC–HRMS method

Yufang Zheng iD, Magnus A B Axelsson iD, Moa Andresen Bergström iD

Journal of Analytical Toxicology · 2025

DOI: 10.1093/jat/bkae087 ↗

Kurz gefasst

Validierungsstudie in der Zeitschrift Journal of Analytical Toxicology, 2025. Verfasst von Yufang Zheng und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Laboratory Medicine, Unilabs, Malarsjukhuset, Eskilstuna 631 88, Sweden. Hauptthemen laut MEDLINE: Saliva, Substance Abuse Detection, Tandem Mass Spectrometry, Illicit Drugs. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 37 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Liquid chromatography–mass spectrometry (LC–MS) methods for detection of multiple drugs of abuse (DoA) in oral fluid (OF) samples are being implemented in many clinical routine laboratories. Therefore, there is a need to develop new multianalyte methods with simple sample pretreatment and short analysis times. The purpose of this work was to validate a method detecting 58 DoA to be used with two different OF sampling kits, the saliva collection system (SCS) from Greiner Bio-One and Quantisal from Immunalysis, using the same sample pretreatment and analytical method. A set of 110 samples collected with the SCS kit was further compared to an high-resolution mass spectrometry (LC–HRMS) method in another laboratory. The method was successfully validated, with precision and accuracy of ≤15% and z-scores of <2 for external controls. Using a sensitive LC–MS–MS instrument, the detection limits were <1 µg/l in neat oral fluid. In the comparative study between the LC–MS–MS and LC–HRMS methods using SCS samples, a good agreement was observed. Discrepancies were limited to lower concentration ranges, attributable to differences in cut-off thresholds between the methods. This work contributes to the development of LC–MS multianalyte methods for OF samples, which are suitable for clinical routine laboratories.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

SalivaSubstance Abuse DetectionTandem Mass SpectrometryIllicit DrugsHumansChromatography, LiquidLimit of DetectionReproducibility of ResultsLiquid Chromatography-Mass Spectrometry

Substanzen

Illicit Drugs

Autor:innen und Institutionen

  • Department of Laboratory Medicine, Unilabs, Malarsjukhuset, Eskilstuna 631 88, Sweden.
  • Region Västra Götaland, Sahlgrenska University Hospital, Laboratory of Clinical Chemistry, Bruna Stråket 16, Gothenburg 413 45, Sweden.
  • Department of Laboratory Medicine, Institute of Biomedicine, Sahlgrenska Academy, University of Gothenburg, Box 100, Gothenburg 405 30, Sweden.

ORCID-Profile: Yufang Zheng, Magnus A B Axelsson, Moa Andresen Bergström

Zitieren

APA

Zheng, Y., Axelsson, M. A. B., & Andresen Bergström, M. (2025). Validation of an LC–MS–MS method for analysis of 58 drugs of abuse in oral fluid and method comparison with an established LC–HRMS method. Journal of Analytical Toxicology. https://doi.org/10.1093/jat/bkae087

Vancouver

Zheng Y, Axelsson MAB, Andresen Bergström M. Validation of an LC–MS–MS method for analysis of 58 drugs of abuse in oral fluid and method comparison with an established LC–HRMS method. Journal of Analytical Toxicology. 2025. doi:10.1093/jat/bkae087

BibTeX
@article{Zheng2025,
  author = {Zheng, Yufang and Axelsson, Magnus A B and Andresen Bergström, Moa},
  title = {Validation of an LC–MS–MS method for analysis of 58 drugs of abuse in oral fluid and method comparison with an established LC–HRMS method},
  journal = {Journal of Analytical Toxicology},
  year = {2025},
  doi = {10.1093/jat/bkae087}
}