Single High-Sensitivity Cardiac Troponin Measurement to Rule-out/Rule-in Myocardial Infarction in the Emergency Department

Anne Marie Dupuy, Yanis Gueche, Anne Sophie Bargnoux, Marion Morena, Bruno Lehodey, François Roubille, Jean Paul Cristol iD

The Journal of Applied Laboratory Medicine · 04.11.2026 · 11 (6) · 1406-1418

DOI: 10.1093/jalm/jfag130 ↗

Kurz gefasst

Fachartikel in der Zeitschrift The Journal of Applied Laboratory Medicine, online erschienen am 31.08.2026. Verfasst von Anne Marie Dupuy und 6 weiteren Autor:innen. Beteiligte Einrichtung: Laboratoire de Biochimie et Hormonologie, CHU Montpellier, Univ. Montpellier 1, Montpellier, France. Hauptthemen laut MEDLINE: Myocardial Infarction, Troponin I, Emergency Service, Hospital, Point-of-Care Testing. Die Arbeit nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: QuidelOrtho Clinical Diagnostics.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Point-of-care (POC) testing and triage liaison physicians could help alleviate the impact of emergency department (ED) overcrowding. The aim of our study was to evaluate the diagnostic performance of the TriageTrue POC assay for the rapid rule-out of myocardial infarction based on a single measurement of high-sensitivity cardiac troponin I (hs-CTnI). The diagnostic performance was compared to measurements obtained with the Vitros 3600 assay, taking into account the cutoffs recommended by the European Society of Cardiology.

Methods

The study compared results of troponin obtained simultaneously on the TriageMeterPro device from whole blood and plasma samples (n = 139) and the hs-cTnI assay on the Vitros 3600 instrument from EDTA plasma samples (n = 225). The diagnostic performance of the 3 methods was compared, including negative predictive value and sensitivity as well as positive predictive value and specificity. Concordance between the methods with disease state samples was then assessed using the Cohen kappa test.

Results

An excellent correlation between plasma and whole blood samples was achieved with the TriageTrue with a coefficient of correlation >0.99 and a bias of +8.2 (±45, with 95% CI 0.4 to 16.0) ng/L. A significant overestimation (around 47%) of the values obtained with the TriageTrue compared with those of the Vitros was also observed. Our results confirmed the acceptable and comparable diagnostic accuracy to the established hs-cTnI-Vitros assay, with a negative predictive value of 100% to rule-out myocardial infarction.

Conclusion

Our results demonstrated that the TriageTrue hs-cTnI whole blood test could serve as an alternative method for ruling out or confirming non-ST-elevation myocardial infarction in the prehospital setting using a single POC troponin measurement.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Myocardial InfarctionTroponin IEmergency Service, HospitalPoint-of-Care TestingHumansBiomarkersSensitivity and SpecificityPredictive Value of TestsPoint-of-Care SystemsFemaleMaleMiddle AgedRapid Diagnostic TestsTriageAged

Substanzen

Troponin IBiomarkers

Autor:innen und Institutionen

  • Laboratoire de Biochimie et Hormonologie, CHU Montpellier, Univ. Montpellier 1, Montpellier, France.
  • PhyMedExp, Université de Montpellier, INSERM, CNRS, Laboratoire de Biochimie et Hormonologie, CHU de Montpellier, France.
  • Département des Urgences, Univ. Montpellier 1, Montpellier, France.
  • Département de Cardiologie, Univ. Montpellier 1, Montpellier, France.

ORCID-Profile: Jean Paul Cristol

Förderung

  • QuidelOrtho Clinical Diagnostics

Zitieren

APA

Dupuy, A. M., Gueche, Y., Bargnoux, A. S., Morena, M., Lehodey, B., Roubille, F., & Cristol, J. P. (2026). Single High-Sensitivity Cardiac Troponin Measurement to Rule-out/Rule-in Myocardial Infarction in the Emergency Department. The Journal of Applied Laboratory Medicine, 11(6), 1406-1418. https://doi.org/10.1093/jalm/jfag130

Vancouver

Dupuy AM, Gueche Y, Bargnoux AS, Morena M, Lehodey B, Roubille F, et al. Single High-Sensitivity Cardiac Troponin Measurement to Rule-out/Rule-in Myocardial Infarction in the Emergency Department. The Journal of Applied Laboratory Medicine. 2026;11(6):1406-1418. doi:10.1093/jalm/jfag130

BibTeX
@article{Dupuy2026,
  author = {Dupuy, Anne Marie and Gueche, Yanis and Bargnoux, Anne Sophie and Morena, Marion and Lehodey, Bruno and Roubille, François and Cristol, Jean Paul},
  title = {Single High-Sensitivity Cardiac Troponin Measurement to Rule-out/Rule-in Myocardial Infarction in the Emergency Department},
  journal = {The Journal of Applied Laboratory Medicine},
  year = {2026},
  volume = {11},
  number = {6},
  pages = {1406--1418},
  doi = {10.1093/jalm/jfag130}
}
Kostenlos starten →