Comparison of Two Ion-Exchange HPLC Methods for HbA1c Quantitation

Darren L Latzig, Paula Zwiefelhofer, Ian L Gunsolus

The Journal of Applied Laboratory Medicine · 04.11.2026 · 11 (6) · 1452-1462

DOI: 10.1093/jalm/jfag112 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift The Journal of Applied Laboratory Medicine, online erschienen am 28.07.2026. Verfasst von Darren L Latzig und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Laboratory Medicine and Pathology, HealthPartners, Bloomington, MN, United States. Hauptthemen laut MEDLINE: Glycated Hemoglobin. Die Arbeit nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Tosoh Bioscience, Inc.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

This study characterized the analytical performance of the Tosoh Automated Glycohemoglobin Analyzer HLC-723GR01 (GR01), the newest-generation hemoglobin A1c (HbA1c) analyzer from Tosoh, relative to the predicate Tosoh HLC-723G8 glycohemoglobin analyzer (G8).

Methods

Performance of the GR01 was assessed through precision, linearity, method comparison, and variant interference studies in whole blood and hemolysate. The G8 was the comparator method for all studies. Variant interference studies included specimens with heterozygous hemoglobin variants C, D, E, and S, as well as elevations of hemoglobin A2. Interference studies for elevated hemoglobin F were also conducted using cord blood.

Results

Total imprecision for the GR01 at HbA1c concentrations ranging from 5.0% to 12.0% (31–108 mmol/mol) was approximately 1% CV, meeting common acceptance limits applied to HbA1c clinical testing and method certification. Linearity studies also met acceptance criteria, with an average of 0.1% bias from the expected HbA1c concentration across the analytical measurement range. All paired GR01 and G8 HbA1c measurements in nonvariant specimens showed relative differences less than 5%; mean differences (in %HbA1c units) were 0.03% in whole blood (SD 0.07%) and 0.02% HbA1c in hemolysate (SD 0.07%). Across all variants, the mean difference between GR01 and G8 measurements was 0.1% (SD 0.25%).

Conclusions

The GR01 showed excellent correlation with the predicate G8, including in specimens containing high concentrations of common heterozygous hemoglobin variants, exceeding variant levels of 30% to 40%. Relative bias between the methods near medical decision points of 6.5% and 8.0% (48 and 64 mmol/mol, respectively) achieved the ±5% criterion used by the National Glycohemoglobin Standardization Program.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Glycated HemoglobinHumansChromatography, High Pressure LiquidReproducibility of ResultsChromatography, Ion Exchange

Substanzen

Glycated Hemoglobinhemoglobin A1c protein, human

Autor:innen und Institutionen

  • Department of Laboratory Medicine and Pathology, HealthPartners, Bloomington, MN, United States.

Förderung

  • Tosoh Bioscience, Inc

Zitieren

APA

Latzig, D. L., Zwiefelhofer, P., & Gunsolus, I. L. (2026). Comparison of Two Ion-Exchange HPLC Methods for HbA1c Quantitation. The Journal of Applied Laboratory Medicine, 11(6), 1452-1462. https://doi.org/10.1093/jalm/jfag112

Vancouver

Latzig DL, Zwiefelhofer P, Gunsolus IL. Comparison of Two Ion-Exchange HPLC Methods for HbA1c Quantitation. The Journal of Applied Laboratory Medicine. 2026;11(6):1452-1462. doi:10.1093/jalm/jfag112

BibTeX
@article{Latzig2026,
  author = {Latzig, Darren L and Zwiefelhofer, Paula and Gunsolus, Ian L},
  title = {Comparison of Two Ion-Exchange HPLC Methods for HbA1c Quantitation},
  journal = {The Journal of Applied Laboratory Medicine},
  year = {2026},
  volume = {11},
  number = {6},
  pages = {1452--1462},
  doi = {10.1093/jalm/jfag112}
}
Kostenlos starten →