Assessment of Intra‐ and Intercapillary Variability in HbA1c Measurement Using the Sebia Capillarys OCTA3 Analyzer
ELECTROPHORESIS · 2026
DOI: 10.1002/elps.70064 ↗Kurz gefasst
Fachartikel in der Zeitschrift ELECTROPHORESIS, 2026. Verfasst von Meriem Belhedi und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Laboratories, Habib Thameur's Hospital, Tunis, Tunisia. Hauptthemen laut MEDLINE: Glycated Hemoglobin. Die Arbeit nennt 35 Literaturangaben.
Abstract (Originalsprache)
Hemoglobin A1c (HbA1c) measurement is crucial for diabetes monitoring. Validation of analytical methods ensures the reliability of results. We verified the analytical performance of HbA1c measurement in each capillary of the Capillarys OCTA3 automated capillary analyzer, following the French accreditation committee guide. For each capillary and each HbA1c level, we evaluated repeatability, intermediate precision, accuracy, robustness, and measurement uncertainty. Results were compared to acceptance limits from French Society of Clinical Biology (SFBC), European Federation of Clinical Chemistry and Laboratory Medicine (EFLM), Réseau d'identification et de Communication en Oncologie et Sénologie (RICOS), and Clinical Laboratory Improvement Amendments (CLIA). A comparison with the high‐performance liquid chromatography (HPLC) method was performed on 37 samples using linear regression and Bland–Altman analysis. Repeatability coefficient of variations (CVs) were less than 1.1% for both levels, consistent with supplier and SFBC (3.8%) thresholds. Precision CVs were 1.24% (normal) and 1.3% (pathological), below SFBC and supplier limits but above the strict EFLM standard. Accuracy showed consistent biases between 0% and 1.02%. Method comparison showed acceptable agreement. Expanded uncertainties were 3.10% (normal) and 2.89% (pathological), exceeding EFLM recommendations (1.8%). Westgard and CLIA sigma indices confirmed satisfactory robustness; EFLM indices were lower for some capillaries, limiting Six Sigma use. The OCTA3 analyzer demonstrated excellent analytical performance. Individual capillary verification yielded satisfactory results. Despite uncertainty exceeding EFLM recommendations and variable robustness, this method is reliable for routine clinical use.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Laboratories, Habib Thameur's Hospital, Tunis, Tunisia.
- Department of Biochemistry, Main Military Hospital of Tunis, Tunis, Tunisia.
- Department of Biochemistry, Béchir Hamza Children's Hospital, Tunis, Tunisia.
ORCID-Profile: Othman Bacha, Afif Ba
Zitieren
APA
Belhedi, M., Lazzem, W., Bacha, O., Ba, A., Rym, D., & Chouaieb, S. (2026). Assessment of Intra‐ and Intercapillary Variability in HbA1c Measurement Using the Sebia Capillarys OCTA3 Analyzer. ELECTROPHORESIS. https://doi.org/10.1002/elps.70064
Vancouver
Belhedi M, Lazzem W, Bacha O, Ba A, Rym D, Chouaieb S. Assessment of Intra‐ and Intercapillary Variability in HbA1c Measurement Using the Sebia Capillarys OCTA3 Analyzer. ELECTROPHORESIS. 2026. doi:10.1002/elps.70064
BibTeX
@article{Belhedi2026,
author = {Belhedi, Meriem and Lazzem, Wiem and Bacha, Othman and Ba, Afif and Rym, Dabboubi and Chouaieb, Sonia},
title = {Assessment of Intra‐ and Intercapillary Variability in HbA1c Measurement Using the Sebia Capillarys OCTA3 Analyzer},
journal = {ELECTROPHORESIS},
year = {2026},
doi = {10.1002/elps.70064}
}
EUVIMED