EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Distinct intracardiac electrogram waveforms with perforation during left bundle branch area pacing implantation

Heli Tolppanen, Valerian Valiton iD, Samuel Stempfel iD, Haran Burri iD

Europace · 03.02.2026 · 28 (2)

DOI: 10.1093/europace/euag010 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Europace, online erschienen am 17.02.2026. Verfasst von Heli Tolppanen und 3 weiteren Autor:innen. Beteiligte Einrichtung: Cardiac Pacing Unit, Department of Cardiology, University Hospital of Geneva, rue Gabrielle Perret Gentil 4, Geneva 1211, Switzerland. Hauptthemen laut MEDLINE: Cardiac Pacing, Artificial, Electrophysiologic Techniques, Cardiac, Heart Injuries, Bundle-Branch Block. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 23 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Finnish Cardiac Society, Finska Läkaresällskapet, Paavo Nurmi Foundation.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Aims

Perforation during left bundle branch area pacing (LBBAP) results in a fall in the current of injury (COI) amplitude in the unipolar unfiltered electrogram (iEGM), but systematic waveform analyses have not been performed. Our aim was to investigate unipolar iEGM waveforms during perforation and to compare them to those at the final lead position. Methods and results The iEGMS of consecutive patients who had perforation during LBBAP implantation were systematically analysed. A total of 92 patients with perforation were included. In the unfiltered channel, sensed COI amplitude was lower with perforation [3.0 (1.5–4.1) mV] than at the final lead position [14.0 (9.2–17.5) mV, P < 0.0001], as was also the case during pacing. Patients with narrow QRS/non-LBBB typically had wide negative (QS) waveforms during sensing (in 67% of cases), whereas those with LBBB/paced rhythm had positive (wide R/RS) morphologies (in 93% of cases). In the former subgroup, a sensed Q or S amplitude > COI amplitude (which is easy to eyeball during lead deployment) had a sensitivity of 86% and a specificity of 93% for diagnosing perforation. Waveforms during macroperforation (with loss of capture, n = 27) differed compared to microperforation (with preserved capture, n = 65), with significantly lower COI amplitudes, more frequent QS morphology, and rarer sharp multiphasic components in the ventriculogram of the filtered channel.

Conclusion

Beyond COI amplitude, additional iEGM waveform parameters may be used to evaluate the presence of LBBAP perforation and should be carefully monitored during lead deployment to improve safety.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Cardiac Pacing, ArtificialElectrophysiologic Techniques, CardiacHeart InjuriesBundle-Branch BlockBundle of HisPacemaker, ArtificialHumansFemaleMaleAgedElectrocardiographyPredictive Value of Tests

Stichwörter der Autor:innen

ComplicationsConduction system pacingCurrent of injuryElectrogramLeft bundle branch area pacingPerforation

Autor:innen und Institutionen

  • Cardiac Pacing Unit, Department of Cardiology, University Hospital of Geneva, rue Gabrielle Perret Gentil 4, Geneva 1211, Switzerland.
  • Heart and Lung Center, Helsinki University Central Hospital, Helsinki, Finland.

ORCID-Profile: Valerian Valiton, Samuel Stempfel, Haran Burri

Förderung

  • Finnish Cardiac Society
  • Finska Läkaresällskapet
  • Paavo Nurmi Foundation

Zitieren

APA

Tolppanen, H., Valiton, V., Stempfel, S., & Burri, H. (2026). Distinct intracardiac electrogram waveforms with perforation during left bundle branch area pacing implantation. Europace, 28(2). https://doi.org/10.1093/europace/euag010

Vancouver

Tolppanen H, Valiton V, Stempfel S, Burri H. Distinct intracardiac electrogram waveforms with perforation during left bundle branch area pacing implantation. Europace. 2026;28(2). doi:10.1093/europace/euag010

BibTeX
@article{Tolppanen2026,
  author = {Tolppanen, Heli and Valiton, Valerian and Stempfel, Samuel and Burri, Haran},
  title = {Distinct intracardiac electrogram waveforms with perforation during left bundle branch area pacing implantation},
  journal = {Europace},
  year = {2026},
  volume = {28},
  number = {2},
  doi = {10.1093/europace/euag010}
}
Kostenlos starten →