Shortened radiofrequency delivery time to optimize efficiency and safety of pulmonary vein isolation with the radiofrequency balloon: insights from the COLLABORATE registry
Europace · 2024
DOI: 10.1093/europace/euae227 ↗Kurz gefasst
Multizentrische Studie in der Zeitschrift Europace, 2024. Verfasst von Alexandre Almorad und 21 weiteren Autor:innen. Beteiligte Einrichtung: Heart Rhythm Management Centre, Postgraduate Program in Cardiac Electrophysiology and Pacing, Universitair Ziekenhuis Brussel-Vrije Universiteit Brus…. Hauptthemen laut MEDLINE: Pulmonary Veins, Registries, Catheter Ablation, Atrial Fibrillation. Die Arbeit nennt 34 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Aims
Previous clinical studies on pulmonary vein isolation (PVI) with a radiofrequency balloon (RFB) reported safe and effective procedures using conventional ablation settings with 20/60 s RF delivery via posterior/anterior (PST/ANT) electrodes. The latest evidence suggests that reducing the application time to 15 s (s) on the posterior wall when facing the oesophageal region is as effective as applying 20 s. To prospectively assess whether reducing RF time on PST/ANT segments to 15/45 s can ensure sufficient quality of lesion metrics and compare the new shortened ablation settings with the conventional one in terms of safety, and effectiveness at 1-year. Methods and results A total of 641 patients from seven European centres were enrolled in a collaborative registry, with 374 in the conventional RF delivery group and 267 in the shortened RF delivery group. Procedural outcomes, lesion metrics, and safety profiles were assessed and compared between the groups. Freedom of any atrial tachycarrythmias at one year was 85.4% and 88.2% in the SHRT and CONV groups, respectively. The shortened RF delivery strategy was associated with significantly shorter procedure times (median 63.5 vs. 96.5 min, P < 0.001) and shortened fluoroscopy exposure (median 10.0 vs. 14.0 min, P < 0.001) compared to conventional delivery. Efficacy metrics, including first-pass isolation rates and time to isolation, were comparable between groups. Shortened RF delivery was associated with a lower incidence of procedural complications (1.4% vs. 5.3%, P = 0.04) and optimized thermal characteristics.
Conclusion
Analyses from the COLLABORATE registry demonstrate that shortening RF energy delivery times to 15/45 s (PST/ANT) during PVI with the RFB resulted in comparable freedom from recurrent atrial tachyarrhythmia compared to conventional delivery times with comparable efficiency and safety.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Heart Rhythm Management Centre, Postgraduate Program in Cardiac Electrophysiology and Pacing, Universitair Ziekenhuis Brussel-Vrije Universiteit Brussel, Laarbeeklaan 101, 1090 Brussels, Belgium.
- Cardiology Department, Hartcentrum Hasselt, Jessa Ziekenhuis, Belgium.
- Cardiology Lab, Gentofte Hospital, University of Copenhagen, Gentofte, Denmark.
- Cardiology Department, CHU Helora Jolimont, Jolimont, Belgium.
- Cardiology Department, Mediterranea Cardiocentro, Naples, Italy.
- Cardiology Department, University of Naples Federico II, Naples, Italy.
ORCID-Profile: Alexandre Almorad, Domenico Giovanni Della Rocca, Alvise Del Monte, Johan Vijgen, Pieter Koopman, René Worck, Arne Johannessen, Antoine de Meester de Ravenstein, Teresa Strisciuglio, Sara Poggi, Giuseppe Stabile, Carmelo La Greca, Joseph Antoine Kheir, Milad El Haddad, Charles Audiat, Roberto Scacciavillani, Luigi Pannone, Carlo de Asmundis
Zitieren
APA
Almorad, A., Rocca, D. G. D., Del Monte, A., Vijgen, J., Koopman, P., Worck, R., Johannessen, A., Lepièce, C., de Ravenstein, A. M., Strisciuglio, T., Poggi, S., Stabile, G., La Greca, C., Kheir, J. A., Jesel-Morel, L., El Haddad, M., Hossein, A., Audiat, C., Scacciavillani, R., …, & Chierchia, G. B. (2024). Shortened radiofrequency delivery time to optimize efficiency and safety of pulmonary vein isolation with the radiofrequency balloon: insights from the COLLABORATE registry. Europace. https://doi.org/10.1093/europace/euae227
Vancouver
Almorad A, Rocca DGD, Del Monte A, Vijgen J, Koopman P, Worck R, et al. Shortened radiofrequency delivery time to optimize efficiency and safety of pulmonary vein isolation with the radiofrequency balloon: insights from the COLLABORATE registry. Europace. 2024. doi:10.1093/europace/euae227
BibTeX
@article{Almorad2024,
author = {Almorad, Alexandre and Rocca, Domenico Giovanni Della and Del Monte, Alvise and Vijgen, Johan and Koopman, Pieter and Worck, René and Johannessen, Arne and Lepièce, Caroline and de Ravenstein, Antoine de Meester and Strisciuglio, Teresa and Poggi, Sara and Stabile, Giuseppe and La Greca, Carmelo and Kheir, Joseph Antoine and Jesel-Morel, Laurence and El Haddad, Milad and Hossein, Amin and Audiat, Charles and Scacciavillani, Roberto and Pannone, Luigi and de Asmundis, Carlo and Chierchia, Gian-Battista},
title = {Shortened radiofrequency delivery time to optimize efficiency and safety of pulmonary vein isolation with the radiofrequency balloon: insights from the COLLABORATE registry},
journal = {Europace},
year = {2024},
doi = {10.1093/europace/euae227}
}
EUVIMED